Figure 1-3: Mount
A. 1
½
-in. x 1 in. reducing bushing
B. 1
½
-in. pipe tee PN 2002011
Straight flow shown
Note
Install sensor at least ten degrees above the horizon.
Table 1-1: Horizontal Pipe Tee (PN 2002011) Pressure/Temperature
Ratings
psig (kPa)
°F (°C)
150 (1136)
150 (65)
128 (984)
160 (71)
102 (805)
170 (77)
80 (653)
180 (82)
57 (494)
200 (93)
48 (432)
210 (99)
November 2020
Quick Start Guide
Quick Start Guide
7