Figure 1-14: Rosemount 3900/3900VP without Preamplifier to
Rosemount 5081 Transmitter Wiring
Table 1-9: Rosemount 3900/3900VP without Preamplifier to Rosemount 5081
Transmitter Wiring
Wire color
Wire function
Terminal block number
Inner drain
No connection
N/A
Green
N/A
Ground
White
RTD return
3
White/red
RTD sense
4
Red
RTD in
5
White/gray
Reference shield
6
Gray
Reference in
7
Blue
Solution ground
8
Clear
mV/pH shield
9
Orange
mV/pH in
10
N/A
N/A
11
N/A
N/A
12