pH sensors with integral smart preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
II 1 G Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Specific Conditions for Safe Use (X):
1. All pH/ORP sensor models with a plastic enclosure or exposed plastic
parts may provide an electrostatic ignition hazard and must only be
cleaned with a damp cloth to avoid the danger of ignition due to
build-up of electrostatic charge.
2. All pH/ORP sensor models with a metallic enclosure may provide a
risk of ignition by impact or friction. Care should be taken during
installation to protect the sensor from the risk.
3. External connections to the sensor must be suitably terminated and
provide a degree of protection of at least IP20.
4. All pH/ORP sensor models are intended to be in contact with the
process fluid and may not meet the 500V r.m.s. test to earth. This
must be taken into consideration at installation.
5.7
International
5.7.1
IECEx Intrinsic Safety
Certificate
IECEx BAS 10.0083X
Standards
IEC 60079-0: 2011, IEC 60079-11: 2011
Markings
pH/ORP sensors with no preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
pH sensors with integral smart preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +60 °C)
ORP sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
pH sensors with integral standard preamplifier fitted
Ex ia IIC T4 Ga (–20 °C to +80 °C)
Ex ia IIC T5 Ga (–20 °C to +40 °C)
Quick Start Guide
November 2020
32
Emerson.com/Rosemount
Содержание Rosemount 3900
Страница 12: ...Figure 1 7 Low Flow Panel PN SQP10077 LQD Quick Start Guide November 2020 12 Emerson com Rosemount ...
Страница 35: ...6 Declaration of Conformity November 2020 Quick Start Guide Quick Start Guide 35 ...
Страница 36: ...Quick Start Guide November 2020 36 Emerson com Rosemount ...
Страница 37: ...7 China RoHS table November 2020 Quick Start Guide Quick Start Guide 37 ...
Страница 38: ...A Intrinsically safe sensor installation drawing FM Quick Start Guide November 2020 38 Emerson com Rosemount ...
Страница 39: ...November 2020 Quick Start Guide Quick Start Guide 39 ...