Note
Transmitters are shipped fully calibrated per request or by the factory
default of full scale (span = upper range limit).
The transmitter will only allow 3–5 percent URL Zero error to be trimmed.
For greater zero errors, compensate for the offset by using the XD_Scaling,
Out_Scaling and Indirect L_Type which are part of the AI Block
Procedure
Guided Setup
1. Navigate to
Configure
→
Guided Setup
.
2. Select
Zero Trim.
The method will execute the zero trim.
Manual Setup
3. Navigate to
Overview
→
Calibration
→
Sensor Trim
.
4. Select
Zero Trim.
The method will execute the zero trim.
Quick Start Guide
June 2020
20
Emerson.com/Rosemount