5
Maintaining and Troubleshooting
5.1
Maintaining the sensor
WARNING!
HIGH PRESSURE
Retraction chamber contents may be under pressure. Reduce the pressure to 0 psig before
opening the retraction chamber.
Failure to reduce pressure may cause a loose part to disengage and cause injury to personnel.
WARNING!
TOXIC LIQUIDS
Be sure the sensor has been cleaned of process liquid before handling.
Generally, the only maintenance required is to keep the opening of the sensor clear of
deposits. Cleaning frequency is best determined by experience.
5.2
Troubleshooting
Resistance vs. Temperature for Temperature Compensation (PT-100 RTD)
Table 5-1:
Temperature
Resistance
10 °C (50 °F)
103.9 Ω
20 °C (68 °F)
107.8 Ω
25 °C (77 °F)
109.7 Ω
30 °C (86 °F)
111.7 Ω
40 °C (104 °F)
115.5 Ω
50 °C (122 °F)
119.4 Ω
Maintaining and Troubleshooting
Rosemount 228 Sensor
29
Содержание Rosemount 228
Страница 1: ...Reference Manual LIQ MAN 228 Rev V April 2018 Rosemount 228 Toroidal Conductivity Sensors ...
Страница 2: ......
Страница 4: ......
Страница 6: ...Contents ii Reference Manual ...
Страница 26: ...Wiring diagram for Rosemount 1056 and 56 Transmitters Figure 3 2 Wiring 20 Reference Manual ...
Страница 34: ...Calibration 28 Reference Manual ...
Страница 36: ...Maintaining and Troubleshooting 30 Reference Manual ...
Страница 38: ...Accessories 32 Reference Manual ...
Страница 40: ...Return of Materials 34 Reference Manual ...
Страница 41: ...Return of Materials Rosemount 228 Sensor 35 ...
Страница 42: ...Return of Materials 36 Reference Manual ...
Страница 43: ...Return of Materials Rosemount 228 Sensor 37 ...