5.3
Choosing Temperature Units and Automatic/Manual
Temperature Compensation
5.3.1
Purpose
Most liquid analytical measurements (except ORP) require temperature compensation.
The Model 1057 performs temperature compensation automatically by applying internal temperature
correction algorithms. Temperature correction can also be turned off. If temperature correction is
off, the Model 1057 uses the temperature entered by the user in all temperature correction
calculations.
5.3.2
Procedure
Follow the menu screens in
Figure 5-1
to select automatic or manual temp compensation, set the
manual reference temperature, and to program temperature units as °C or °F.
5.4
Configuring and Ranging The Current Outputs
5.4.1
Purpose
The Model 1057 accepts inputs from three sensors and has four analog current outputs. Ranging the
outputs means assigning values to the low (0 or 4 mA) and high (20 mA) outputs. This section
provides a guide for configuring and ranging the outputs. Always configure the outputs first.
5.4.2
Definitions
1. Current ouputs - The analyzer provides a continuous output current (4-20 mA or 0-20 mA)
directly proportional to the process variable or temperature. The low and high current outputs can be
set to any value.
2. Assigning outputs - Assign a measurement to outputs 1, 2, 3, or 4.
3. Dampen - Output dampening smooths out noisy readings. It also increases the response time of
the output. Output dampening does not affect the response time of the display.
4. Mode - The current output can be made directly proportional to the displayed value (linear mode)
or directly proportional to the common logarithm of the displayed value (log mode).
S1: 1.234µS/cm 123.4°C
S2: 12.34pH 123.4°C
S3:123.4µS/cm 123.4°C
Temperature
Units: °C
S1 Temp Comp: Auto
S2 Temp Comp: Auto
S3 Temp Comp: Auto
Figure 5-1 Choosing Temperature Units and Manual Auto Temp Compensation
Section 5: Programming the Analyzer - Basics
22
Instruction Manual
LIQ-MAN-1057
Section 5: Programming the Analyzer - Basics
April 2017
Содержание Rosemount 1057
Страница 1: ...Instruction Manual LIQ MAN 1057 Rev D April 2017 Rosemount 1057 Three Input Intelligent Analyzer ...
Страница 2: ...A Essential Instructions Read this page before proceeding You ...
Страница 4: ...ii ...
Страница 6: ...Quick Start Guide Figure A Quick Start Guide Quick Start Guide iv ...
Страница 7: ...Quick Reference Guide Figure B Model 1057 Menu Tree Quick Reference Guide v ...