2
Mounting
Topics covered in this chapter:
•
•
•
•
Rotate the electronics on the meter (optional)
•
Rotate the display on the transmitter (optional)
2.1
Prepare the installation
Before you can install the meter and fit the thermal insulation, verify that the system is
stable and leak free.
1.
Fit a blanking compression nut to the meter mounting, pressurize, and flush the
system.
2.
Isolate the system, depressurize, and remove the blanking compression nut.
2.2
Mount the meter
There are three ways to mount the meter.
Related information
Mount with a flow-through chamber
Mount with an inline viscosity retrofit kit
Mount with a capillary viscosity retrofit kit
2.2.1
Mount with a flow-through chamber
Flow-through chambers are manufactured by Micro Motion, and are available with either
of the following:
•
Welded ends or compression fittings that connect into the process pipelines
•
1- inch, 2-inch, or 3-inch inlet and outlet pipes
Important
Do not alter the length of the inlet and outlet pipes. Pipe alterations can adversely affect the fitting
temperature response and stability.
Prerequisites
Verify the following conditions:
Mounting
Installation Manual