6.2.3
Configure
Digital Communications Fault Action
ProLink III
Device Tools
>
Configuration
>
Fault Processing
Overview
Digital Communications Fault Action
specifies the values that will be reported via digital
communications if the device encounters an internal fault condition.
Procedure
Set
Digital Communications Fault Action
as desired.
The default setting is
None
.
Options for
Digital Communications Fault Action
Options for
Digital Communications Fault Action
Table 6-3:
Label
Description
ProLink III
Upscale
Upscale
•
Process variable values indicate that the value is great-
er than the upper sensor limit.
•
Totalizers stop incrementing.
Downscale
Downscale
•
Process variable values indicate that the value is lower
than the lower sensor limit.
•
Totalizers stop incrementing.
Zero
IntZero-All 0
•
Flow rate variables go to the value that represents a
flow rate of 0 (zero).
•
Density is reported as
0
.
•
Temperature is reported as
0 °C
, or the equivalent if
other units are used (e.g.,
32 °F
.
•
Drive gain is reported as measured.
•
Totalizers stop incrementing.
None
None
(default)
•
All process variables are reported as measured.
•
Totalizers increment if they are running.
Integrate the meter with the control system
Configuration and Use Manual
49
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...