Alert num-
ber
Alert title
Possible cause
Recommended actions
A020
Calibration Factors
Missing
Some calibration factors have not
been entered or are incorrect.
•
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
•
Verify the setting of the
Sensor Type
pa-
rameter.
•
If
Sensor Type
=
Curved Tube
, ensure that
no parameters specific to
Straight Tube
have been set.
A021
Transmitter/Sensor/
Software Mismatch
The configured board type does not
match the physical board, or the
configured sensor type does not
match the physical sensor.
•
Verify all of the characterization or cali-
bration parameters. See the sensor tag
or the calibration sheet for your meter.
•
Ensure that the correct board is instal-
led.
•
Verify the setting of the
Sensor Type
pa-
rameter.
•
If
Sensor Type
=
Curved Tube
, ensure that
no parameters specific to
Straight Tube
have been set.
A027
Security Breach
The transmitter has been switched
from secure mode to unsecure
mode. In unsecured mode, the
transmitter configuration can be
changed.
•
Secure the transmitter by setting the
CT switch to ON.
Note that additional procedures may be
required to return the device to secured
mode.
A030
Incorrect Board Type The loaded software is not compati-
ble with the programmed board
type.
•
Contact Micro Motion.
A033
Insufficient Pickoff
Signal
The signal from the sensor pick-
off(s) is insufficient. This suggests
that the sensor tubes or vibrating
elements are not vibrating. This
alert often occurs in conjunction
with Alert 102.
•
Check for air in the flow tubes, tubes
not filled, foreign material in the tubes,
coating in the tubes, or other process
problems.
•
Check for foreign material in the proc-
ess gas or fluid, coating, or other proc-
ess problems.
•
Check for fluid separation by monitor-
ing the density value and comparing
the results against expected density val-
ues.
•
Ensure that the sensor orientation is ap-
propriate for your application. Settling
from a two-phase or three-phase fluid
can cause this alert even if the flow
tubes are full.
Troubleshooting
Configuration and Use Manual
73
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...