Alert num-
ber
Alert title
Possible cause
Recommended actions
A011
Zero Calibration
Failed: Low
Many possible causes, such as too
much flow, especially reverse flow,
through the sensor during a calibra-
tion procedure, or a zero result that
is too low. This alert is accompanied
by A010.
This alert will not clear until you cy-
cle power to the meter.
•
Verify that there is no flow through the
sensor, cycle power to the meter, then
retry the procedure.
A012
Zero Calibration
Failed: High
Many possible causes, such as too
much flow, especially forward flow,
through the sensor during a calibra-
tion procedure, or a zero result that
is too high. This alert is accompa-
nied by A010.
This alert will not clear until you cy-
cle power to the meter.
•
Verify that there is no flow through the
sensor, cycle power to the meter, then
retry the procedure.
A013
Zero Calibration
Failed: Unstable
There was too much process insta-
bility during the calibration proce-
dure.
This alert will not clear until you cy-
cle power to the meter.
•
Remove or reduce sources of electro-
mechanical noise (e.g., pumps, vibra-
tion, pipe stress), cycle power to the
meter, then retry the procedure.
A014
Transmitter Failure
Many possible causes.
•
Ensure that all wiring compartment
covers are installed correctly.
•
Ensure that all transmitter wiring meets
specifications and that all cable shields
are properly terminated.
•
Check the grounding of all compo-
nents.
•
Evaluate the environment for sources of
high electromagnetic interference
(EMI) and relocate the transmitter or
wiring as necessary.
•
Contact Micro Motion.
A016
Sensor Temperature
(RTD) Failure
The value computed for the resist-
ance of the line RTD is outside lim-
its.
•
Check your process conditions against
the values reported by the device.
•
Check the wiring between the sensor
and the transmitter.
•
Verify that internal wiring is secure and
that there are no internal electrical
problems.
•
Contact Micro Motion.
Troubleshooting
72
Micro Motion
®
LNG Series Meters
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...