Base Time Unit
is the existing time unit that the special unit will be based on.
3.
Calculate
Mass Flow Conversion Factor
as follows:
a. x base units = y special units
b.
Mass Flow Conversion Factor
= x ÷ y
The original mass flow rate value is divided by this value.
4.
Enter
Mass Flow Conversion Factor
.
5.
Set
Mass Flow Label
to the name you want to use for the mass flow unit.
6.
Set
Mass Total Label
to the name you want to use for the mass total and mass
inventory unit.
The special measurement unit is stored in the transmitter. You can configure the
transmitter to use the special measurement unit at any time. You must use ProLink III to
select the special measurement unit.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1.
Set
Base Mass Unit
to
Pounds
(lb).
2.
Set
Base Time Unit
to
Seconds
(sec).
3.
Calculate
Mass Flow Conversion Factor
:
a. 1 lb/sec = 16 oz/sec
b.
Mass Flow Conversion Factor
= 1 ÷ 16 = 0.0625
4.
Set
Mass Flow Conversion Factor
to
0.0625
.
5.
Set
Mass Flow Label
to
oz/sec
.
6.
Set
Mass Total Label
to
oz
.
4.1.2
Configure
Flow Damping
ProLink III
Device Tools
>
Configuration
>
Process Measurement
>
Flow
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement.
Damping
Value
specifies the time period (in seconds) over which the transmitter will spread changes
in the process variable. At the end of the interval, the internal value will reflect 63% of the
change in the actual measured value.
Procedure
Set
Flow Damping
to the value you want to use.
The default value is 0.64 seconds. The range is 0 to 60 seconds.
Configure process measurement
Configuration and Use Manual
19
Содержание Micro Motion LNG Series
Страница 1: ...Configuration and Use Manual MMI 20032809 Rev AA February 2017 Micro Motion LNG Series Meters ...
Страница 6: ...Contents iv Micro Motion LNG Series Meters ...
Страница 8: ...Getting Started 2 Micro Motion LNG Series Meters ...
Страница 12: ...Before you begin 6 Micro Motion LNG Series Meters ...
Страница 18: ...Quick start 12 Micro Motion LNG Series Meters ...
Страница 20: ...Configuration and commissioning 14 Micro Motion LNG Series Meters ...
Страница 50: ...Configure device options and preferences 44 Micro Motion LNG Series Meters ...
Страница 56: ...Integrate the meter with the control system 50 Micro Motion LNG Series Meters ...
Страница 58: ...Complete the configuration 52 Micro Motion LNG Series Meters ...
Страница 60: ...Operations maintenance and troubleshooting 54 Micro Motion LNG Series Meters ...
Страница 73: ...D1 and D2 density calibration using ProLink III Figure 9 1 Measurement support Configuration and Use Manual 67 ...
Страница 74: ...Measurement support 68 Micro Motion LNG Series Meters ...
Страница 94: ...Using ProLink III with the transmitter 88 Micro Motion LNG Series Meters ...
Страница 98: ...Default values and ranges 92 Micro Motion LNG Series Meters ...
Страница 99: ...Default values and ranges Configuration and Use Manual 93 ...