Note the following:
•
If your transmitter does not have a display, you must use ProLink II or the Communicator.
•
The Communicator does not provide a loop test for the discrete input.
•
The mA reading does not need to be exact. You will correct differences when you trim the mA
output. See Section 5.4.
•
If you are connected to the transmitter via HART/Bell 202, the HART/Bell 202 signal will
affect the reading. Disconnect from the transmitter before reading the output, then reconnect
and resume the loop test after taking the reading. If you are using any other protocol, this is not
required.