Configuration and Use Manual
139
Troubleshooting
Meas
ure
m
ent P
e
rf
ormance
Defaults
T
rou
b
le
s
ho
oti
n
g
Compensa
tion
A009
Transmitter
Initializing/Warming
Up
Transmitter in power-up
mode.
• Allow the flowmeter to warm up (approximately 30
seconds). The error should disappear once the
flowmeter is ready for normal operation.
• If alarm does not clear, make sure that the sensor is
completely full or completely empty.
• Check the sensor circuitry. See Section 11.25.
Transmitter
Initializing/Warming
Up
A010
Calibration Failure
Mechanical zero: The
resulting zero was greater
than 3
μ
s.
Temperature/Density Cals:
many possible causes.
• If alarm appears during a transmitter zero, ensure
that there is no flow through the sensor, then retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the
flowmeter to operation.
Calibration Failure
A011
Excess Calibration
Correction, Zero
too Low
See A10
• Ensure that there is no flow through the sensor, then
retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the
flowmeter to operation.
Zero Too Low
A012
Excess Calibration
Correction, Zero
too High
See A10
• Ensure that there is no flow through the sensor, then
retry.
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the
flowmeter to operation.
Zero Too High
A013
Process too Noisy
to Perform Auto
Zero
See A10.
• Remove or reduce sources of electromechanical
noise, then retry. Sources of noise include:
- Mechanical pumps
- Pipe stress at sensor
- Electrical interference
- Vibration effects from nearby machinery
• Cycle power to the flowmeter, then retry.
• If appropriate, restore the factory zero to return the
flowmeter to operation.
Zero Too Noisy
A014
Transmitter Failed
Many possible causes.
• Cycle power to the flowmeter.
• The transmitter might need service. Contact Micro
Motion. See Section 11.3.
Transmitter Failed
A016
Line RTD
Temperature
Out-Of-Range
The value computed for
the resistance of the Line
RTD is outside limits
• Check the sensor RTD circuitry. See Section 11.25.
• Verify that process temperature is within range of
sensor and transmitter.
• If the problem persists, contact Micro Motion. See
Line RTD
Temperature
Out-of-Range
A017
Meter RTD
Temperature
Out-Of-Range
The value computed for
the resistance of the
Meter/Case RTD is
outside limits
• Check the sensor RTD circuitry. See Section 11.25.
• Verify that process temperature is within range of
sensor and transmitter.
• If the problem persists, contact Micro Motion. See
• Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
Meter RTD
Temperature
Out-of-Range
A020
Calibration Factors
Unentered
The flow calibration factor
and/or K1 has not been
entered since the last
master reset.
• Check the characterization. Specifically, verify the
FCF and K1 values. See Section 6.2.
• If the problem persists, contact Micro Motion. See
Calibration Factors
Unentered
(FlowCal)
Table 11-4
Status alarms and remedies
continued
Alarm
code
Communicator
Cause
Suggested remedy
ProLink II
Содержание Micro-Motion 2400S
Страница 8: ...vi Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 20: ...12 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 30: ...22 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 42: ...34 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 66: ...58 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 76: ...68 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 138: ...130 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 190: ...182 Micro Motion Model 2400S Transmitters with Analog Outputs ...
Страница 197: ......