4.3.2
Wire the Ethernet I/O using M12-terminated cables
(optional)
Prerequisites
Obtain one D-coded M12-terminated Ethernet cable.
Procedure
1. Attach the M12-terminated Ethernet cable to the Ethernet I/O connector on the
1600 transmitter.
Figure 4-11: M12-terminated cables to the Ethernet I/O
2. Attach the other cable end using the pinouts described in the following table.
Table 4-2: M12 Ethernet I/O pinouts
Pin
identification
Wire color
Outputs on RJ45
Signal name
Pin 1
Orange/White
Pin1
TDP1/RDP2
Pin 2
Green/White
Pin 3
RDP1/TDP2
Pin 3
Orange
Pin 2
TDN1/RDN2
Pin 4
Green
Pin 6
RDN1/RDN2
Wiring the channels