100-412-196 REV. 06
51
4
Auto Reset
In OFF position the alarm/error must be reset either
by the front panel Reset key or through the external
9-pin connector. The controls will not function and
no parameters can be changed, the cycle result
when the alarm/error occurred will be shown until
the Reset signal has been given.
In ON position a Reset signal is not required. The
Start signal can be given directly after the alarm/
error has occurred. After the alarm/error has
occurred the operator can access all functions of the
LPX system. The overload will need to be cleared
before making any changes.
0 (OFF) Default
1 (ON)
5
Trigger Beeper
The beeper will sound when ultrasonics are
activated.
0 (OFF)
1 (ON) Default
6
Error Beeper
The beeper will sound if an error is encountered.
0 (OFF)
1 (ON) Default
7
Alarm Beeper
The beeper will sound if an alarm is activated.
0 (OFF)
1 (ON) Default
8
Configuration Lock
In the ON position system settings are locked.
Access to ultrasonic cycle parameter modification;
system configuration registers; and saving/loading
cycle configurations is no longer permitted.
In the off position access is unrestricted to all
parameters, system settings, and cycle
configurations.
NOTICE
To turn off Configuration Lock, power down the
power supply, then press and hold down the Mode/
Configuration key while turning on the unit to access
the register settings.
0 (OFF) Default
1 (ON)
9
Seek @ Power Up
In the OFF position, the seek @ power up will not
occur.
In the ON position, the seek @ power up will occur.
When the system performs a Seek function the
ultrasonic stack is run at low amplitude to tune to
the ultrasonic converter's operating frequency.
0 (OFF)
1 (ON) Default
Table 6.3
Register Settings
Register
Description
Parameters
Содержание Branson LPX
Страница 4: ...iv 100 412 196 REV 06 ...
Страница 6: ...vi 100 412 196 REV 06 ...
Страница 8: ...viii 100 412 196 REV 06 ...
Страница 10: ...x 100 412 196 REV 06 Table 7 2 System Trouble Analysis Chart 73 Table 7 3 Alarms Errors 76 ...
Страница 24: ...14 100 412 196 REV 06 ...
Страница 33: ...100 412 196 REV 06 23 Chapter 3 Delivery and Handling 3 1 Delivery and Handling 24 ...
Страница 38: ...28 100 412 196 REV 06 4 1 4 Declaration of Conformity Figure 4 1 Declaration of Conformity ...
Страница 76: ...66 100 412 196 REV 06 ...
Страница 85: ...100 412 196 REV 06 75 Figure 7 2 LPX Power Supply Interconnect Diagram ...