Online View
[73]
Click on
Channel OK
to expand the object, and get more information about the converter
state.
Table 9-1: Flags Channel OK
Flag
Meaning
Action
Channel OK
No fault detected.
---
Sensor or converter fault
detected.
Check sensor and converter including
wiring and connections
Sensor is too close to the
measuring target.
Check the distance between sensor
and measuring target and readjust the
sensor. See
Temperature
The temperature, measured
by the internal sensor is
within the OK range of -35°C
to 85°C.
---
The temperature, measured
by the internal temperature
sensor, is out of the OK
range of -35°C to 85°C
Take appropriate measures to reduce
the environmental temperature.
Voltage output
This graphic object displays the DC part of converter output voltage. This voltage is
proportional to the distance between sensor and measuring object.
Figure 9-4: Voltage output
Calibration state
This graphic object displays the state of the calibration.
Figure 9-5: Calibration state – expanded object
•
A fault- free calibration is indicated with a checkmark within a green solid circle .
Содержание AMS EZ1000
Страница 39: ... 35 6 7 Control drawings ...
Страница 40: ... 36 ...