Configuration
[65]
CAUTION!
Machine Studio will automatically establish an online connection to the EZ 1000 as
soon as there is a physical connection through the USB port.
2.
Click
Send & close
in the ribbon command bar to send the configuration to the EZ
1000. The configuration editor automatically closes after the sending process.
A successfully sent configuration will be indicated by a message in the upper right
corner of the software window. This message window will automatically disappear.
Otherwise close it by clicking on the cross.
The converter is ready to use when the sensor is connected, the converter output
voltage is within the OK range, and the green
Ok
LED on the converter front shows a
steady light.
Note
After sending a configuration the configuration becomes active immediately – this is
indicated by Machine Studio and a steady green LED on the converter front. The output of the
EZ 1000 is activated 25 seconds after the sending process.
8.6.2
Reload a configuration
Once an online connection has been established, the configuration of the EZ 1000 is
automatically loaded to Machine Studio. Click
Reload
in the ribbon command bar if the
configuration and the calibration data must be loaded again.
8.7
Offline (Easy) calibration
Any work on the system may impair machine protection.
The EZ 1000 converter can be calibrated on EZ 108x sensors with 5 m or 10 m sensor cable
without using Machine Studio, provided that the material of the measuring object is
42CrMo4.
Note
The offline calibration requires an EZ 1000 with factory default settings (delivery state). If uncertain
about the calibration state, use the online view of Machine Studio (see
) to check the
current calibration type of the converter –
Auto
,
Easy
, or
Manual
. Factory default settings are also
indicated by a continuously triple flashing of the LED (see
Reset an already configured EZ 1000 converter to the factory default settings by using Machine
Studio (see
) before starting the offline calibration procedure.
An already offline (easy) calibrated EZ 1000 can be offline calibrated again without resetting the
converter to factory default.