8. “Set-Point”, pH (on/off)
pH reading values can be set to operate the pH pump using 2 set-points into
On/Off mode or Proportional (%).
On/Off mode set the instrument to operate using two set values that enable or disable the pH pump. To use this
mode move cursor on “Working Mode”. Press the wheel and select it.
ON/OFF mode while dosing ALKALI
Set pH value at 7.00 OFF and 6.90 ON.
Instrument will leave the pH pump active until reading value will increase up to 7.00pH.
At 7.00pH the pH pump will be disabled until reading value will decrease under 6.90pH.
Waiting Time: to let pump operate at pulses per minutes add one or more minute (1pulse every xx minutes). Otherwise (00 minutes) pump will
operate at set stroke settings (see page 22).
6.90 7.00
ON
OFF