ESD3000-System
34/66
6.5.1.2 Point exploration with air discharge
The points on an EUT to which the discharges should be applied can be discovered by first doing an
exploratory test by running at a repetition rate of approximately 20 discharges per second or more.
With the ESD3000 and the ESD3000DM1 it can be arranged as following:
Select the desired discharge voltage in air discharge mode.
charging voltage
polarity positive and voltage in kV
voltage level selected L2 = 4 kV
press ENTER button and the
display change contact to air
trigger mode manual
Press the „RUN“ button and hold the trigger button pressed on the main body
When the finger button is held
down the discharge tip is supplied
with d.c. The discharges will occur
with a repetition of approximately
20 Hz.
6.5.1.3 Putting out of operation
As showed above the
„RUN“
button must be pressed until the message is shown
„power down“.
Hold this finger button
pressed