ESD3000-System
16/66
2.3 Precautionary measure during use
The ESD3000
generates high voltages. The energy content of the 30 kV modules is high
and can be dangerous with improper use. It is wise to observe the following rules
•
Never touch the EUT when a test is in operation.
•
Touch no connectors of connection cable or tips when a EMC test is in operation.
•
The high voltage of the ESD3000 and the power on the EUT must turned off before a manipulation on
the EUT is carried out.
•
For all tests the 2 m ground cable must be connected to ground.
2.4 Electromagnetic Compatibility
The outputs of the ESD3000 and the discharge to the EUT will emit disturbances.
Please consider the national emission rules.
The Test System ESD3000 should not be operated near sensitive measuring and
control systems.
The ESD3000 fulfils the following immunity requirements
•
Electrostatic discharge
Level 4 (8 kV)
(IEC 61000-4-2)
2.5 The manual is an integral part of the equipment. Refer to the manual.
This manual is an integral part of the ESD3000. The safety rules and precautions in the manual must
be observed. EMC PARTNER and their representatives are not responsible for damage to persons
and equipment by not observance the safety rules and precautions in the manual.
2.6 No operation of ESD3000 without ground wire
The ESD3000 must be operated with a ground wire attached to the battery case
Содержание ESD3000
Страница 14: ...ESD3000 System 14 66 1 6 Accessories dimensions 1 6 1 Included articles dimensions 1 6 2 Standard accessories...
Страница 24: ...ESD3000 System 24 66...
Страница 40: ...ESD3000 System 40 66 replace the batteries by pull it out piece by piece...
Страница 42: ...ESD3000 System 42 66...
Страница 44: ...ESD3000 System 44 66...
Страница 50: ...ESD3000 System 50 66...
Страница 54: ...ESD3000 System 54 66...
Страница 60: ...ESD3000 System 60 66...
Страница 62: ...ESD3000 System 62 66...
Страница 64: ...ESD3000 System 64 66...
Страница 66: ...ESD3000 System 66 66...