EN
Range of SonoTT
TM
Clamp-On Transducers and Their Flow Measurement Range
ID
Type
Qmin
3
Qmax
4
Tube Size
(Inner diameter (ID) x Wall Thickness (WT))
13482
CT 5,7 mm
100 ml/min
±1000 ml/min
5.7 mm outer diameter
13479
CT 6,8 mm
100 ml/min
±6000 ml/min
6,8 mm outer diameter
13481
CT 3/16" x 1/16”
100 ml/min
±6000 ml/min
3/16“ x 1/16”
13478
CT 1/4" x 3/32”
120 ml/min
±8 000 ml/min
1/4“ x 3/32”
13475
CT 1/4" x 1/16”
120 ml/min
±8 000 ml/min
1/4“ x 1/16”
13476
CT 3/8" x 3/32”
150 ml/min
±10 000 ml/min
3/8” x 3/32”
13480
CT 3/8" x 1/16”
150 ml/min
±10 000 ml/min
3/8” x 1/16”
13477
CT 1/2" x 3/32”
300 ml/min
±20 000 ml/min
1/2“ x 3/32”
3,4
Please note
: While a measurement is possible for the flow ranges stated above, the accuracy is only defi
-
ned for the flow range between the Qmin and the Qmax as stated under “Accuracy in Combination with the
SonoTT
TM
Ultrasonic FlowComputer".
Please note
: If the medium within the tube moves in the opposite direction of the arrow on the lid, the flow value is
displayed as negative value (e.g. -250 ml⁄min instead of 250 ml/min). The accuracy and the flow range stated above is
the same for positive as well as negative flow values.
Please specify calibration parameters at time of ordering, i.e. media, temperature and tubing type.
SonoTT
TM
Clamp-On Transducer
Page 24
D162-700 SonoTT
TM
Clamp-On Transducer - Operating Instructions V3.0
Copyright em-tec GmbH