Volume Conversion Device EK260
30
Elster GmbH
The EK260 itself assumes a linear characteristic and after the entry of
Prog
(see
below) only determines the coefficients
Eq1p
and
Eq2p
. The third coefficient
Eq3p
remains unchanged. The standard value for
Eq3p
is "0".
p1Adj Adjustment value 1 for pressure
p2Adj Adjustment value 2 for pressure
Prog
Accept pressure adjustment
These values are used for the adjustment of the pr essure measurement circuit, i.e.
for the internal computation of the equation coefficients for the pressure (see
above).
The adjustment takes place in three steps:
1. Apply measurement pressure 1 (= reference value 1) to the pressure se nsor
and enter as
p1Adj
.
2. Apply measurement pressure 2 (= reference value 2) to the pressure sensor
and enter as
p2Adj
.
3. Enter
Prog
= "1" so that the EK260 calculates the equation coefficients.
After applying the measurement pressure, you should either wait about one minute
each time before entering the adjustment or press <ENTER> number of times
during the display of the pressure measurement
p.Mes
(see below) until the
displayed value is stable.
Approx. 0.4
ž
pMax
and approx. 0.9
ž
pMax
should be selected as adjustment
values.
p.Amb Air pressure fixed value
p.Mes Pressure measurement
p.Abs Absolute pressure measurement
p.Abs
is the sum of
p.Amb
and
p.Mes
:
p.Abs
=
p.Amb
+
p.Mes
.
"0" should be entered when using an absolute pressure sentors and the ambient
pressure when using a gauge pressure sensor.
p.Mes
is the absolute or gauge pressure depending on the pressure sensor.
If the measured pressure
p.Abs
is within the alarm limits
pMin
and
pMax
(see
above), then it is used as the pressure
p
for correction:
p
=
p.Abs
.
p.MP
Ø
Measurement period mean pressure
p.MP Ø
is the averaged value of all pressure measurements within the current
measurement period.
At the end of each measurement period
p.MP Ø
is saved in the measurement
period archive (
→
3.6).
p.Mon max Month's maximum pressure
p.Mon min Month's minimum pressure
p.Mon max
is the highest,
p.Mon min
the lowest of all pressure measurements
within the current month.
By entering <ENTER> you can branch to the relevant submenu where the
corresponding time stamp is displayed.
The maxima and minima of the last 15 months can be interrogated in the monthly
archive 2 (
→
3.6).
Содержание 83 462 140
Страница 117: ...Volume Conversion Device EK260 Elster GmbH 117 A 2 Approval for Ex Zone 1 ...
Страница 118: ...Volume Conversion Device EK260 118 Elster GmbH ...
Страница 119: ...Volume Conversion Device EK260 Elster GmbH 119 ...
Страница 120: ...Volume Conversion Device EK260 120 Elster GmbH ...
Страница 121: ...Volume Conversion Device EK260 Elster GmbH 121 ...