EnCal 3000 Quad– Hardware Manual
22/06/2017
4/52
Safety Information
Electrical Safety
The EnCal 3000 Quad is a gas analyser designed to be installed in hazardous areas. For this purpose it is
certified according to:
ATEX II 2 G Ex d IIB T4 Gb and IECEx Ex d IIB T4 Gb
In compliance with:
IECEx:
IEC 60079-0: 2011 Sixth edition
IEC 60079-1:2014 seventh edition
ATEX:
EN 60079-0: 2012
EN 60079-1: 2014
Detailed information on this certificate can be found in APPENDIX
The operation and maintenance of such equipment should only be performed by qualified personnel.
Fuse replacement must only be performed by Elster GmbH authorized personnel.
The following basic rules must be observed in all circumstances:
The flameproof enclosure may not be opened when an explosive gas atmosphere may
be present. The process pressure shall be limited to 2 MPa to ensure that the pressure
rise inside the flameproof enclosure remains below 10 kPa. The functional pressure
must be lower see chapter 3 and section 5.1.4 containing the specifications. Before any
start-up, verify all connections of the unit for tight sealing. For information on the
dimensions of the flameproof joints the manufacturer shall be contacted.
WARNING! ELECTROSTATIC HAZARD!
Because of the potential electrostatic charging of the paint layer, the housings should
only be cleaned by using a moist cloth to prevent charging
BOTH CASES MUST BE EARTHED!
Important disclaimer:
If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
Installation instructions mentioned in this manual are intended for information only. The installation of this
equipment must be conform to any national, local, or company codes applicable to the location. Elster-
Instromet assumes no responsibility for compliance with these requirements. It is suggested that a review of
the codes be made prior to installation.