EnCal 3000 Quad– Hardware Manual
22/06/2017
33/52
Location
Outdoor installation, close to sample point. Only sun shade required
Mounting
Wall mounted
Gas connections
Swagelok 1/8” or on request 3mm
Electrical connections
Power Supply
M20/M25 cable gland for armoured cable Ø 11-17 mm / 17-26 mm
Data Communication M20/M25 cable gland for armoured cable Ø 11-17 mm / 17-26 mm
Installation
Helium / Argon
Quality 5.0 or better
Optional 2 bottles with automatic change system
Supply pressure 5.5 barg
Flow ± 4 ml/min per column (max. 16 ml/min with four columns)
Calibration gas
Composition preferably close to pipe line composition
Quality 2.0 or higher
(with a maximal uncertainty of 1% relative deviation for each component)
Supply pressure 1 to 4 barg nominal
Pressure peak protection up to 4 barg
Flow ± 30 ml/min