24 - EN
liberoCONFIG
LI6002Eg
CONFIGURATION SOFTWARE
3
The end of logging is automatically determined by the start time and the selected
logging interval.
Start Options
Start Mode
Determines the oper-
ating condition of the
LIBERO Tx after it has
been configured.
• Start immediately
Data recording starts as soon as the LIBERO Tx is configured.
• Start after pressing the TRANSIT-button
Data recording starts after the TRANSIT-button has been pressed for 2 seconds.
Initial state: Transit / Arrived Initial State
Depending on the transportation cycles, you may switch several times between the two
modes:
• Start in Transit mode
The LIBERO Tx starts recording with consideration of the alarm settings.
• Start in Arrived mode
The LIBERO Tx starts recording without consideration of the alarm settings.
This mode is used for transport monitoring where the products are ready for shipment,
but the transport starts sometimes later e.g. to respect cool down phase of packaging.
Just before the transport starts, press the TRANSIT-button to activate the alarm moni-
toring.
Both modes could be delayed.
Содержание LIBERO Tx
Страница 1: ...LIBERO Tx PDF Logger Operation Manual...
Страница 38: ...38 EN liberoCONFIG LI6002Eg CONFIGURATION SOFTWARE 3...
Страница 41: ...EN 41 liberoCONFIG LI6002Eg CONFIGURATION SOFTWARE 3...
Страница 79: ...EN 79 liberoCONFIG LI6002Eg MISCELLANEOUS 7...