23.1. Measurement with automatic temperature compensation
In order to make measurement with automatic temperature compensation:
- connect the oxygen sensor and the temperature probe to the O2 and t
connectors respectively (Pic. 2), the symbol will appear;
- immerse the oxygen sensor and the temperature probe in the measured
solution;
- switch the meter on with the
button;
- choose the measuring function with the
button;
- choose the unit according to the section 21.1;
- in case of accurate measurements in mg/l enter the salinity value
(section 21.2);
- check or simulate flow of the measured solution;
- wait about 1 min until the reading stabilises and check it.
23.2. Measurement with manual temperature compensation
In order to make measurement with manual temperature compensation:
- disconnect the temperature probe;
- switch the meter on with the
button;
- choose the measuring function with the
button;
- choose the unit according to the section 21.1;
- in case of accurate measurements in mg/l enter the salinity value
(section 21.2);
- immerse the oxygen sensor in the measured solution;
- measure the temperature of the solution with a laboratory thermometer;
- with the
,
buttons enter the temperature value of the measured
solution into the lower row of the LCD (pressing of both
,
buttons
simultaneously sets the temperature to 20
o
C);
- check or simulate flow of the measured solution;
- check the reading when it stabilises.
In case of measurement series it is possible to adjust the temperature
compensation each time the temperature of the solution changes.
Uwaga:
in case of measurements in solutions with low salinity, check
according to the section 21.2 if the introduced salinity value is
equal 0.00 g/l.
Содержание CX-401
Страница 1: ...WATERPROOF MULTIFUNCTION METER CX 401 USER S MANUAL...
Страница 2: ......
Страница 3: ...USER S MANUAL WATERPROOF MULTIFUNCTION METER CX 401 Before use please read the instructions carefully...
Страница 4: ......
Страница 7: ...1 I INTRODUCTION...
Страница 8: ...2 User s manual for CX 401 multifunction meter v6 0x...
Страница 15: ...9 Pic 2...
Страница 25: ...19 II pH MEASUREMENT...
Страница 26: ...20 User s manual for CX 401 multifunction meter v6 0x...
Страница 44: ...38 User s manual for CX 401 multifunction meter v6 0x...
Страница 45: ...39 III CONDUCTIVITY AND SALINITY MEASUREMENT...
Страница 46: ...40 User s manual for CX 401 multifunction meter v6 0x...
Страница 71: ...65 IV OXYGEN CONCENTRATION MEASUREMENT...
Страница 72: ...66 User s manual for CX 401 multifunction meter v6 0x...
Страница 86: ...80 User s manual for CX 401 multifunction meter v6 0x...
Страница 87: ...81 V ATMOSPHERIC PRESSURE MEASUREMENT...
Страница 88: ...82 User s manual for CX 401 multifunction meter v6 0x...
Страница 90: ...84 User s manual for CX 401 multifunction meter v6 0x...
Страница 91: ...85 VI REDOX POTENTIAL AND TEMPERATURE MEASUREMENT...
Страница 92: ...86 User s manual for CX 401 multifunction meter v6 0x...
Страница 96: ...90 User s manual for CX 401 multifunction meter v6 0x...
Страница 97: ...91 VII OTHER...
Страница 98: ...92 User s manual for CX 401 multifunction meter v6 0x...
Страница 116: ......
Страница 117: ......
Страница 118: ......
Страница 119: ......