6) Specifications
GENERAL SPECIFICATIONS
Display : 3-3/4 digits 4000 counts LCD display(s)
Update Rate : 3 per second nominal
Polarity : Automatic
Low Battery : Below approx. 2.4V
Operating Temperature : 0
O
C to 40
O
C
Relative Humidity : Maximum relative humidity 80% for temperature up to 31°C decreasing
linearly to 50% relative humidity at 40
O
C
Altitude : Operating below 2000m
Storage Temperature : -20
O
C to 60
O
C, < 80% R.H. (with battery removed)
Temperature Coefficient : nominal 0.15 x (specified accuracy)/
O
C @(0
O
C -18
O
C or 28
O
C -
40
O
C), or otherwise specified
Sensing : Average sensing for 3036s and 126s. True RMS for 3038s
Safety : Meets IEC/UL/EN61010-1 ed. 3.0, IEC/EN61010-2-032 ed. 3.0, IEC/EN61010-2-033
ed. 1.0, IEC/EN/UL61010-031 ed. 1.1 & CAN/CSA-c22.2 no. 61010-1-12 ed. 3.0
Measurement Category : III 600 Volts ac & dc
Pollution degree : 2
E.M.C. : Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges: Total Accuracy = Specified Ac 45 digits
Performance above 3V/m is not specified
Overload Protections :
ACA Clamp-on jaws : AC 400A rms continuous
+/
µ
A & COM terminals (all functions) : 600VDC/VAC rms
Power Supply : 1.5V alkaline button batteries (IEC/JIS LR44, GPA76, or V13GA). Two
batteries for 3036 and 3038, and four batteries for 126s
Power Consumption : 2.5 mA typical
APO Timing : Idle for 30 minutes
APO consumption : 5
µ
A TYPICAL
Dimension : L190mm X W63mm X H32mm
Weight : 187 gm approx
Jaw opening & Conductor diameter : 26mm max
Accessories : Test leads (pair), batteries installed, user's manual, & soft carrying pouch.
Banana plug type-K bead probe Bkp60 x 1 (BM126s only)
Optional Accessories : Banana pins to type-K socket plug adapter Bkb32 (for Elma126s only)
Special features : 30ms max hold; data hold; relative zero mode (models 3036s and 3038s);
simultaneous A+V, A+
Ω
, A+
o
C (Elma 126); slim jaws; light weight
Electrical Specifications
Accuracy is ±(% reading number of digits) or otherwise specified, at 23
O
C ±5
O
C & less
than 75% R.H.
1)
Model Elma 3038 True RMS accuracy of ACV is specified from 5 % to 100 % of range, and
ACA is specified from 10 % to 100 % of range, or otherwise specified. Maximum Crest Factor <
1.75 : 1 at full scale & < 3.5 : 1 at half scale, and with frequency component within the specified
frequency bandwidth for non-sinusoidal waveforms