AM852 Veterinary Anaesthesia machine User’ Manual
22
5. Installation and Replacing
5.1Install absorber
Warning:
Please comply with prescript of security defense in point:
·
Avoid skin or eyes to contact with material in absorber. In case skin or eyes contact
with it, rinse affected parts with rinsing immediately and take medical measures
·
Do not replace absorber during ventilation.
·
Check the color of absorbent after every case. The color of absorbent could resume
former color during nonuse. Refer to absorbent label to obtain detail about color
change.
·
Completely dry absorbent will release CO when contacting with anesthetic. Please
replace absorbent in time taking security into account.
5.2 Replace absorbent
·
Gradual color change of calx sodica in absorber indicates CO
2
absorption, but the
color change of calx sodica is a approximate show. Utilize a CO
2
detector to determine
whether to replace absorbent or not.
·
Discard the absorbent once it change color. Calx sodica could resume former color
when deposited for several hours, which will mislead users.
1. Unload absorber: unscrew screw, and loosen absorber so as to take it down.
2. Discard scrap calx sodica, and then fill new absorbent.
3. Fix absorber: place absorber on the former position, and screw screw.
·
There are two absorber, and each may accommodate 1000mL absorbent.
Connect gas source
Note:
Only medical gas source can be used. Other types of gas source may contain water,
oil or other garbage.
Connect one end of input pipe of O
2
, N2O and air with inlet tie-in of pipeline and screw
down. Connect another end to corresponding gas source.