 User Manual TrackSense LyoPro – MAN 7 

 

 

This is required for a complete 100R system setup using buffer vial: 

 

  

 

 

1.

 

31201000

 – LyoPro data logger with thermocouple sensor 

 

2.

 

31203001

 – 100R LoggerNest 

 

3.

 

31203062

 – 100R LoggerClip 

 

4.

 

Buffer Vial 100R w. rubber stopper 

 

5.

 

3120303004

 – 100R BufferClip 

 

6.

 

Vial 100R w/o rubber stopper 

 

7.

 

31203003

 – 100R LyoStopper 

 

11.

 

Firmware Update  

11.1.

 

Reader Station and LyoPro Loggers  

Purpose:

  

Firmware updater for reader station and logger. Install the TS LyoPro firmware 

updater located on the provided USB key. 

 

 

Procedure for installing the firmware upgrading tools for LyoPro Reader Station 

and Loggers: 

 

 

1.

 

Double click the 

TSLyoProFirmwareUpdateSetup_xxxxyyzz.msi  

program. 

 

Содержание TrackSense LyoPro

Страница 1: ...TrackSense LyoPro Version BETA Page 1 of 46 User Manual User Manual TrackSense LyoPro MAN 7 TrackSense LyoPro User Manual...

Страница 2: ...Installation Requirements 12 4 Activating ValSuite 13 4 1 Online Activation The PC is Connected to the Internet 13 4 1 1 Manual Activation The PC is Not Connected to the Internet 14 4 1 2 Releasing th...

Страница 3: ...Versions Reference Numbers 30 9 2 LoggerNest 30 9 2 1 Standard Versions Reference Numbers 31 9 3 LoggerClip 31 9 3 1 Standard Versions Reference Numbers 32 9 4 BufferClip 32 9 4 1 Standard Versions Re...

Страница 4: ...ng and stability studies The customer needs to verify the temperature range of the equipment being used before placing it into an application 1 2 General DO NOT use the equipment for any application o...

Страница 5: ...to the insulation due to contact with hot surfaces cuts or abrasions Always replace the power cord with an approved cord of the correct rating and type If you have questions contact an Ellab Authoriz...

Страница 6: ...mum pressure under which the TrackSense LyoPro may be operated is 4 Bar ABS Permitted range of ambient temperature 80 C Tamb 60 C short term up to 121 C 1 4 4 Important Notice Regarding Ellab LyoPro E...

Страница 7: ...ins licence exempt transmitter s receiver s that comply with Innovation Science and Economic Development Canada s licence exempt RSS s Operation is subject to the following two conditions 1 This devic...

Страница 8: ...8 of 46 User Manual User Manual TrackSense LyoPro MAN 7 When contacting Ellab for support please have the following information available Model Number Serial Number Voltage Complete description of th...

Страница 9: ...llab reader station is intended to be used at room temperature The station should not be exposed to temperatures below 5 C or above 50 C Do not immerse the reader station in water under any circumstan...

Страница 10: ...ader station can be used at a time 3 2 Communication Setup All TrackSense LyoPro 10 slot Reader Stations can be connected to PC via USB or LAN ethernet 3 2 1 USB connection Connect the USB cable to th...

Страница 11: ...1 Connect the 12 Vac power supply and ethernet cable to the reader station 2 Connect the reader station to the PC using the USB cable 3 Read and note the reader station IP address located under the st...

Страница 12: ...ws7 SP1 Windows 8 1 Windows10 Windows Server 2008 Service Pack 2 Windows server 2008 R2 and Windows server 2012 R2 Software support 64 Bits on platform Windows7 SP1 Windows 8 1 Windows10 Windows Serve...

Страница 13: ...ped according to GAMP principles and is available in multiple languages for Windows 7 8 and 10 32 64 Bit For further information please consult the specific ValSuite Software user manual Note The firs...

Страница 14: ...lect the Manual activation option if the PC is not connected to the internet 2 Send an e mail to Ellab with the computer ID Example of computer ID F0DEF14E64E41 3 In the ValSuiteactivation field write...

Страница 15: ...Start Logging Data with TrackSense LyoPro Starting TrackSense LyoPro Loggers 1 Ensure that the battery has enough capacity to cover the expected session length If the battery is new activate it separa...

Страница 16: ...USB default or Ethernet connection and click OK 5 Select LyoPro Start Logger A New Session window will then appear Enter a name for the session mandatory session text optional and click OK A window wi...

Страница 17: ...can be programmed Choose Settings at the bottom of the screen The Logger Settings window will appear Define Sample Rate and optionally Programmed Start Stop conditions and activate Primary Control Ch...

Страница 18: ...mouse when it is placed on the graph A Select Data Viewer compare data sets B Select Interval Data Viewer numerically display data C Select Add Label place special information directly in the graph D...

Страница 19: ...contacts verify that the data logger is not leaking Lightly clean the area using a cotton swab and alcohol then change the o rings Any data logger with recurring problems should be returned to Ellab f...

Страница 20: ...tivated battery according to the indication on the housing pole downwards Put the lid back and tighten the spanner by hand The battery should be placed in separate activator for 60 minutes before inse...

Страница 21: ...battery change Apply a light coat of grease to the o ring Be careful and keep the grease away from any battery contacts Logger O Ring Ordering code O Ring 66151901 Battery compartment o ring 35900053...

Страница 22: ...sed in 30 to 80 C applications In order to get the most accurate voltage reading cool the batteries in a freezer for 1 hour Immediately insert the battery into the activator unit and use a voltmeter t...

Страница 23: ...ro Batteries The TrackSenseLyoPro data loggers have a standby battery usage that will empty the battery in approx 12 months even if the data logger is not used It is therefore recommended to remove th...

Страница 24: ...Pro Battery TrackSenseLyoPro data loggers do not use enough power to activate the battery by itself Therefore the activator unit must be used New batteries must be activated initially for exactly 1 ho...

Страница 25: ...able If received the ETS firmware will also be updated The software will by default automatically warn users 12 months after the last calibration date which should be considered a recommendation for t...

Страница 26: ...sensitive drift can therefore occur if not handled properly Only use cables USB provided by Ellab The ETS system must be properly grounded Never use the ETS outside the measuring range Any ETS with r...

Страница 27: ...tware or it can be setup in connection with the ValSuite software Setup Stand Alone Unit w Display 1 Connect the ETS using a USB cable to the ETS display 2 Connect the display to the power supply Setu...

Страница 28: ...r users with ETS firmware versions older than 2 0 8 2 3 Ellab Dry Blocks Ordering code Model Power Measuring Range 20000155 PTC 155A 115 to 230 VAC 50 60Hz 25 to 155 C 20000425 PTC 425A 115 to 230 VAC...

Страница 29: ...oftware includes three different methods for calibrating data loggers Manual Calibration Semi Automatic only for temperature and Full Automatic only for temperature 8 3 1 Manual Calibration First ente...

Страница 30: ...e Description 31203023 31203043 31203063 31203083 31203103 31203153 31203203 31203253 31203303 31203503 31203003 LyoStopper for 2R vials LyoStopper for 4R vials LyoStopper for 6R vials LyoStopper for...

Страница 31: ...ls LoggerNest for 15R vials Logger Nest for 20R vials LoggerNest for 25R vials LoggerNest for 30R vials LoggerNest for 50R vials LoggerNest for 100R vials 9 3 LoggerClip The LoggerClip is the connecti...

Страница 32: ...als ConnectionClip for 15R vials ConnectionClip for 20R vials Connection Clip for 25R vials ConnectionClip for 30R vials Connection Clip for 50R vials Connection Clip for 100R vials 9 4 BufferClip As...

Страница 33: ...ufferConnectionClip for 10R vials BufferConnectionClip for 15R vials BufferConnectionClip for 20R vials BufferConnectionClip for 25R vials BufferConnectionClip for 30R vials BufferConnectionClip for 5...

Страница 34: ...gerNest 3 31203042 4R Logger Clip 4 Buffer Vial 4R w rubber stopper 5 31203044 4R Buffer Clip 6 Vial 4R w o rubber stopper 7 31203023 4R LyoStopper This is required for a complete 6R system setup usin...

Страница 35: ...p 4 Buffer Vial 100R w rubber stopper 5 3120303004 100R BufferClip 6 Vial 100R w o rubber stopper 7 31203003 100R LyoStopper 11 Firmware Update 11 1 Reader Station and LyoPro Loggers Purpose Firmware...

Страница 36: ...TrackSense LyoPro Version BETA Page 36 of 46 User Manual User Manual TrackSense LyoPro MAN 7 2 Select Next 3 Select a folder in the Change Directory if necessary and click select Next...

Страница 37: ...1 Procedure for Updating the LyoPro Logger and Reader Station Firmware WARNING Before updating users must check the current battery power since updating loggers will use a lot of battery power If the...

Страница 38: ...oPro MAN 7 1 Open the firmware updater which is placed under Programme x86 by default and select Ellab TS LyoPro Firmware Update 2 Select the Loggers button and push Scan 3 Push Select Deselect to tog...

Страница 39: ...est firmware is check marked or a complete list of available versions 6 Highlight the requested FW version and push Select Firmware 7 The firmware updater program will proceed through 4 stages write f...

Страница 40: ...Manual TrackSense LyoPro MAN 7 11 1 2 Procedure for Updating the LyoPro Reader Station firmware 1 Open the firmware updater which is placed under Programme x86 by default and select Ellab TS LyoPro Fi...

Страница 41: ...ill show available firmware versions to choose from 5 Highlight and push Select FW 6 The firmware update program will go through multiple cycles updating the interface and main board inside the Reader...

Страница 42: ...ive for pricing and details 12 1 TrackSense LyoPro Equipment Warranty Ellab offers an industry leading 2 year warranty on the LyoPro logger The thermocouple sensor is not covered neither is physical d...

Страница 43: ...g Material PEEK Diameter 18 mm each compartment Height 32 mm Weight 27 Gram Memory Capacity 100 000 data points 32 511 samples Minimum Sample Rate 1 Second Maximum Sample Rate 24 Hours Time Accuracy 5...

Страница 44: ...on If the loggers are too hot or too cold to be held by hand they are not to be placed in the reader station When I read my logger it gave me no data What can cause this Option 1 Broken Sensor If the...

Страница 45: ...ro MAN 7 Can I use my computer for other jobs while running a TrackSense LyoPro session Yes you can completely close your computer down if necessary Upon reopening the ValSuite software the software w...

Страница 46: ...TrackSense LyoPro Version BETA Page 46 of 46 User Manual User Manual TrackSense LyoPro MAN 7 15 Changelog Edition Section Change Date Author 1 0 New New 2020 09 21 JER TWF...

Отзывы: