6.1.5.
Procedure for Cleaning TrackSense
®
LyoPro Data Loggers
If any corrosion is present near the battery contacts, verify that no leaking is
occurring. Lightly clean the contacts with a cotton swap and alcohol. Change
the o-ring if necessary. Let the area dry before placing the battery in the data
logger.
6.1.6.
Procedure for Changing the Data Logger O-Ring
The TrackSense LyoPro data logger contains one o-ring, which is located at the
battery compartment lid. The o-ring should be inspected and if needed
replaced with every battery change. Apply a light coat of grease to the o-ring.
Be careful and keep the grease away from any battery contacts.
Logger O-Ring:
Ordering code
O-Ring
66151901
Battery compartment o-ring
35900053
O-ring grease
6.1.7.
Extended Warranty
Ellab offers an extended warranty on all new TrackSense LyoPro data loggers.
The extension is from 2 to 5 years. The warranty needs to be purchased
together with the data logger. Please contact your local Ellab sales
representative for pricing.
6.1.8.
Thermocouple Sensor
6.1.8.1.
Proper Use
•
Never use the sensor outside the measuring range
•
Only use the special screwdriver provided by Ellab to remove
sensor from the data logger. Never use any other tools as this
can damage the screw terminals
•
Any data logger/sensor with recurring problems should be
returned to Ellab for service