TrackSense
®
LyoPro
Version
BETA
Page 17 of 46
User Manual
User Manual TrackSense LyoPro – MAN 7
7.
If an immediate start (unprogrammed) is required, press
Start
8.
The slots will turn green and begin showing the first live data. The data loggers
are now active and can be placed in the process. Press
Close
9.
Alternatively, the data loggers can be programmed: Choose
Settings
at the
bottom of the screen. The Logger Settings window will appear. Define
Sample
Rate
and optionally Programmed Start/Stop conditions and activate
Primary
Control
. Choose
Select All
Apply
Close
10.
After the settings are setup, select
Start
. When all selected slots turn green the
data loggers are programmed and will start logging data. Select
Close
11.
Go to
LyoPro
Logger List
. This will showcase all active/programmed data
loggers. Review the details to confirm that they were programmed correctly