45
SECTION 3
CONTROLLING AND CUSTOMIZING STAINER FUNCTIONS
3.2 Recording Specimen and Reagent Information
Scanning Reagent Bottles with the Barcode Reader
1
From System Setup select
QC/Maintenance Tracking.
2
Select
Enable Reagent Tracking
for each desired reagent (A, B, C, D, E)
.
3
Press
Back
to return to the System Setup menu.
4
Press
Reagents
to access the Reagent Information screen.
5
Select the desired Reagent (A, B, C, D, E) and press
Change
.
6
Scan the barcode of each enabled reagent bottle.
7
Verify that the barcode appears on the Scan and Load Slides menu.