
31
SECTION 2
SETUP AND PREPARATION FOR OPERATION
2.2 Preparing the Stainer for Operation
The Clean Cycle
NOTE:
The Clean cycle uses 20 mL of alcohol divided equally among nozzles C and E to
clean the carousel and stainer bowl after staining. Water is sprayed from the D and B
nozzles and Reagent A is sprayed from the A nozzle. Pressing Standby/Ready performs
the same function.
1
Place an empty carousel in the instrument and close the lid.
CAUTION:
Never place any carousel loaded with specimens in the instrument for a Clean
cycle (including placing the instrument in standby mode). Specimens will be
damaged if they contact reagents sprayed from the nozzles when you press
Clean or Standby.
2
Press
Clean
.
NOTE:
Pressing Stop during the Clean cycle causes the Incomplete Clean
message to be displayed. Press Clean to complete the interrupted cycle.
3
Open the lid and remove the carousel when the Clean cycle is complete.
4
Spray the interior of the bowl with 70 to 100% methanol or ethanol. Wipe
the stainer bowl dry with paper towels.
NOTE:
Perform the Storing the Instrument procedure (Section 5.2), if the
instrument will remain idle for more than 1 week.
NOTE:
At the end of a Clean cycle, cleaning reagents (alcohol or water) remain
in the nozzles. These are primed out of the nozzles during the first steps of a
stain cycle.
Performing Tests
You must perform the Pattern Test and Volume Test before using the instrument. See
Section 6, Nozzle Maintenance and Performance.