Product application, installation and service manual
ELIS PLZE
Ň
a. s.
Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025
Page 9 of 44
ELIS PLZE
Ň
a. s., Lu
č
ní 15, P. O. BOX 126, 304 26 Plze
ň
, Czech Republic, tel.: +420/377 517 711, fax: +420/377 517 722
Es 90 337 K/c
The threshold flow rate (Q
NEC
) is the minimum flow rate at which the meter starts to indicate and process the fluid
flow parameters. At the manufacturer’s plant, Q
NEC
is set at a value corresponding to the flow velocity of 20mm/s. On
customer’s request , the threshold flow rate can be set at any value within the range of Q
NEC
= 0.1
¸
25% Q
4
.
The maximum permitted error in fluid volume measurements at flow rates between Q
1
(including) and Q
2
(excluding) is: 5% irrespective of the fluid temperature.
The maximum permitted error in fluid volume measurements at flow rates between Q
2
(including) and Q
4
(including) is: 1% for the fluid temperature not exceeding 50°C, and
3% for the fluid temperature greater than 50°C.
Piping I.D.
DN 200 ÷ DN 1200
Rated pressure
PN 10, on request PN 16 or PN 25 for piping sizes DN 200 ÷ DN 500
Measurement precision
± 1% for flow rate Q > Q
2
(see Table 2) and fluid temperature up to 50°C (SE 4015)
± 0,5% for flow rate Q > Q
2
and fluid temperature up to 50°C (SE 4025)
Fluid temperature
0 ÷ +150°
С
Ambient temperature
0 ÷ +50°
С
Ambient relative humidity not exceeding 80%
Storage temperature
-10 ÷ +70°C
Visual output via
two-line 16-character alphanumeric LC display unit
Power source
90 ÷ 260 V, 50/60 Hz
Back-up power source
Li battery 3 V (lifetime 5 years)
Power requirement
6VA
Line fuse
T 250mA, 250V
Protection against electric
shock
according
to
standard
Č
SN 332000-4-
41
automatic disconnection from power source in the TN – S network
Protect. class; electronic
unit
IP 65
Protect. class; sensor,
probes
IP 54 (on request IP 68)
Outputs (insulated by
means
impulse output, 0.1 ÷ 10,000 l/imp, impulse length 50ms
of optocouplers
frequency output 0 ÷ 1,000Hz (corresponding to flow rates 0 ÷ Q
4
)
relay output 24VAC/ 0.1A
RS 485 line
Optional accessories
insulated current output 0/4 ÷ 20mA (corresponding to flow rates 0 ÷ Q
4
)
mass flow rate information
flow-rate measurements in both directions, flow direction indication
extended range of fluid temperature (up to +180°C)
sensor protection class IP 68
drinking-water meter version
sensor flanges according to alternative standards (ANSI, JIS)
Table 3 SE 4015/ SE 4025 flow meter specifications