6.2.2.7 Definition of limit values 

 

 LIMIT VALUE 

 

 

Here  the  operator  can  set  altogether  11  (limit) values  of  parameters.  The  detailed  description  of  the  procedures 

concerned is shown in Fig. 24. Select the desired parameter by push-button T1 and confirm the selection by T4. The 
name  of  the  parameter  and  the  associated  measuring  unit  will  then  appear  on  the  first  line  of  the  display  and  the 
second line will show the previously defined limit value (with the exception of the date and time). The limit value unit 
shall always be the same as that selected for data display. For example, if the volume flow rate is displayed in litres per 
second, the limit value of volume flow rate  shall also  be  defined in l/s. If the mass flow rate measurement mode is 
selected and the data are displayed in metric tons, the impulse number shall also be defined in t. 
 

Upon  depressing push-button T2, the  previously  set limit value  will disappear from the  second line and 0  will be 

displayed in the first digit position. Use push-buttons T3 and T2 to  set the digital value and T1 to insert the  division 
signs (a comma in the position of a decimal point, dot in the date and colon in the case of time). 
The day in the week information is to be set as follows: 
0 – Sunday 
1 – Monday 
2 – Tuesday 
3 – Wednesday 
4 – Thursday 
5 – Friday 
6 – Saturday  
 

A figure entered may have up to seven digits. The date and time data shall include initial zeroes; e.g. the date of 3 

July, 2001  shall be recorded as  03.07.01 and the time 7 minutes past 9 a.m. as 09:07:00. Confirm the selection by 
depressing push-button T4. In the case of a commercial (invoicing) meter, neither q

s

, impulse number or sensitivity (low 

flow cutoff) can be reset by the user as these settings are reserved to the duly authorised testing authority. Therefore, 
for commercial meters, these parameters will not appear on the list of limit values to be reset.  
 
 
 
 

List of parameters (limit values, date and time) 

Qmax 

  

   

Maximum (overload) flow rate q

s

 in the given measuring units 

ICIS     

   

Impulse number, defining fluid volume or mass (in selected units) per one impulse at the impulse 
output 

Date    

   

The actual calendar date 

Day of the week The actual day of the week 
Time of the day The actual time of the day 
Low flow cutoff   

The flow rate level, in per cent of q

s

, below which the meter will display and at its outputs indicate 

zero flow rate 

Vol. flow limit    

Maximum volume flow rate level; where  a binary output is  associated  with this  parameter, it  will 
indicate values exceeding this limit 

Mass flow limit   

Maximum mass flow rate; where a binary output is associated with this parameter, it will indicate 
values exceeding this limit 

Volume limit     

Maximum  aggregate  volume;  where  a  binary  output  is  associated  with  this  parameter,  it  will 
indicate values exceeding this limit 

Mass limit 

   

Maximum aggregate mass; where a binary output is associated with this 

parameter,  it  will 

indicate values exceeding this limit 

Temp. limit Maximum temperature; where a binary output is associated with this parameter, it 
 will indicate values exceeding this limit value. 

 
Comment: The limit values for all the above parameters shall be given in units selected using the procedure described 
in  section  6.2.2.4  above.  Should  new  parameter  units  be  selected,  the  limit values  need  be  re-defined  accordingly; 
otherwise the meter function would be incorrect. 

Содержание SONOELIS SE 4015

Страница 1: ...anual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 1 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c U...

Страница 2: ...1 3 Display of measured data 13 6 1 2 Review of the measured quantities 13 6 1 3 Review of the measured quantity units 13 6 1 4 Unit conversion table 13 6 2 Keyboard control functions 13 6 2 1 Data d...

Страница 3: ...ow Meter SONOELIS SE 4015 and SE 4025 Page 3 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c 8 TESTING 13 9 ORDERING 13 10 PACK...

Страница 4: ...the ultrasonic probes the ultrasonic ray is sent in turns in and against the fluid flow direction 3 TECHNICAL DESCRIPTION 3 1 General information The SONOELIS SE 4015 SE 4025 ultrasonic flow meter is...

Страница 5: ...the fluid flow in both directions and indicate the actual fluid flow direction On request the meter sensor can be supplied in an IP 68 housing 3 2 Design specifications 3 2 1 Sensor dimensions L O D D...

Страница 6: ...ted to the piping the main pipe section and two pipe branches holding the ultrasonic probes see Fig 4 In the standard sensor version the body is designed for operating pressure PN 10 made of high qual...

Страница 7: ...ets The grommets are intended for tight fitting of cables of external diameter 6 to 8mm PG 9 and 4 to 6mm PG 7 At the bottom of the box there is also an earthing bolt Both the front panel and the term...

Страница 8: ...5 78 75 100 156 3 200 Q1 Minimum flow rate G min 44 68 8 88 1 110 138 173 220 275 347 440 688 881 m3 h 16 25 32 40 50 63 80 100 126 160 250 320 Q2 Transient flow rate G min 70 4 110 141 176 220 277 35...

Страница 9: ...for flow rate Q Q2 see Table 2 and fluid temperature up to 50 C SE 4015 0 5 for flow rate Q Q2 and fluid temperature up to 50 C SE 4025 Fluid temperature 0 150 Ambient temperature 0 50 Ambient relati...

Страница 10: ...nsor are 5DN and 3DN respectively see Fig 7 FLOW DIRECTION 3 DN 5 DN 5 DN 3 DN FLOW DIRECTION Fig 7 Minimum lengths of straight piping sections Fig 8 Minimum length of straight piping section at the i...

Страница 11: ...ons where there is a piping outlet into open reservoirs or tanks anywhere near see Fig 12 FLOW DIRECTION FLOW DIRECTION Fig 12 Examples of incorrect sensor placement Another factor that may influence...

Страница 12: ...Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c 3 0 3 0 PROBE 1 PROBE 2 Fig 13 Permitted range of one beam sensor rotation with respect to its long...

Страница 13: ...atic and mechanical stresses by means of suitable installation boxes All cables shall be lead outside thermal insulation layers on piping if any Connections to the Pt 100 temperature sensor current ou...

Страница 14: ...ecifications on the flange circumference Unless required otherwise the sensor shall be supplied with flanges according to standard SN EN 1092 1 the alternative solutions are ANSI or JIS flanges 5 2 3...

Страница 15: ...puts and the RS 485 communication interface The X4 CANON 9 connector serves the purpose of connecting the RS 485 interface used in the meter calibration servicing and in company operational setting Th...

Страница 16: ...tion on the display unit Combination of switch positions Measurement Instantaneous flow rate Measurement Instantaneous velocity Programming EEPROM programming Servicing Service The servicing mode is r...

Страница 17: ...temperature range Rated pressure PN Fluid flow direction Protection class information 5 2 5 Sealing of technological meters The SONOELIS flow meters are provided with clamping and stick on factory and...

Страница 18: ...ed to show the volume flow rate in m3 hod or the mass flow rate in metric tons per hour and the second display line to display the total volume in m3 or the total mass in metric tons alternatively wit...

Страница 19: ...s l kg l hour kg hour 1 000 bbl 1 000 tons l min kg min bbl ton l s kg s 1 000 ft3 lb bbl hour tons hour ft3 bbl min tons min 1 000 gal bbl s tons s gal ft3 hour lb hour ft3 min lb min ft3 s lb s gal...

Страница 20: ...articular operating conditions at the user s plant The push buttons T1 to T4 are provided with the following graphic symbols T1 T2 T3 T4 The push button control of the meter is shown in a schematic di...

Страница 21: ...is determined by the manufacturer with respect to the meter application Provided the flow meter has been set for measurements in both directions of flow the sign before the reading indicates the flow...

Страница 22: ...flow in the reverse direction 6 2 1 11 Temperature The temperature reading is only available if a thermometer is included in the meter configuration The reading sensitivity is 0 1 C 6 2 1 12 Density...

Страница 23: ...ncorrect password the system will request a new password entry After three consecutive entries of incorrect passwords the system will switch over to the data display mode and will not permit further e...

Страница 24: ...ed language Upon selecting the language confirm the setting by depressing push button T4 The message on the display will inform the operator of completion of the parameter setting in the newly selecte...

Страница 25: ...be recorded as 03 07 01 and the time 7 minutes past 9 a m as 09 07 00 Confirm the selection by depressing push button T4 In the case of a commercial invoicing meter neither qs impulse number or sensi...

Страница 26: ...ion However great care should be taken in employing this function First of all the zero flow rate condition shall be ensured make sure that the closing valve in the piping is not leaking Only then the...

Страница 27: ...appear on the display LIQUID DOES NOT FLOW Check the zero flow rate condition and depress and release S1 The test will continue by checking whether the passage route for the ultrasonic ray in one dire...

Страница 28: ...e meter needs be put out of service and either sent for repair to the manufacturing plant or a service technician be asked to come and repair the meter on site Should a fault be indicated at the ultra...

Страница 29: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 29 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Страница 30: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 30 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Страница 31: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 31 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Страница 32: ...lation and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 32 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 51...

Страница 33: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 33 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Страница 34: ...on and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 34 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 72...

Страница 35: ...ncluding natural disaster A requirement for warranty repair shall be communicated to the manufacturer by a notice in writing by e mail fax or registered letter Should the manufacturer find the warrant...

Страница 36: ...015 system optional assembly and testing fixtures this manual a statement on the product compliance and delivery note 12 WARRANTY CONDITIONS Unless agreed otherwise between the manufacturer and custom...

Страница 37: ...ORDER NUMBER S E 4 0 5 Sensor design one beam sensor 1 double beam sensor 2 SPECIFICATION Dimension 200 800 0 1 of sensor 250 1250 0 2 DN nominal 300 1600 0 3 flow rate Q3 350 2000 0 4 400 2500 0 5 4...

Страница 38: ...3 0 1 for remote 5 0 2 version m 7 0 3 10 0 4 15 0 5 20 0 6 30 0 7 40 0 8 50 0 9 60 special cabel 1 0 70 special cabel 1 1 80 special cabel 1 2 90 special cabel 1 3 100 special cabel 1 4 non standard...

Страница 39: ...Q4 4 20 mA 3 non standard 0 Binary not required 0 1 output fluid flow direction upstream switch on 0 2 fluid flow direction upstream switch off 0 3 total fluid volume level on exceeding switch on 0 4...

Страница 40: ...l min 0 6 l s 0 7 bbl h 0 8 bbl min 0 9 bbl s 1 0 ft3 h 1 1 ft3 min 1 2 ft3 s 1 3 gal h 1 4 gal min 1 5 gal s 1 6 Q4 1 7 non standard 0 0 Units of volume is not pictured 1 volume m 3 2 m 3 10 3 3 l 4...

Страница 41: ...ver lich sud ls 3 O of flow sud sud ss 4 M meter lich lich ll 5 U Group 000 0 0 0 N 001 0 0 1 I 002 0 0 2 C A T I 255 2 5 5 O Address 000 0 0 0 N 001 0 0 1 002 0 0 2 255 2 5 5 CALIBRATION Calibration...

Страница 42: ...47 48 49 50 51 52 ORDER NUMBER DELIVERY TERMS Number of 1 piece 0 0 1 pieces 2 pieces 0 0 2 3 pieces 0 0 3 999 pieces 9 9 9 1000 pieces or more 0 0 0 Packing no packing 1 standard 2 export 3 non stan...

Страница 43: ...tallation and service manual ELIS PLZE a s Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025 Page 43 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377...

Страница 44: ...SE 4025 Page 44 of 44 ELIS PLZE a s Lu n 15 P O BOX 126 304 26 Plze Czech Republic tel 420 377 517 711 fax 420 377 517 722 Es 90 337 K c Manufacturer s address ELIS PLZE a s Lu n 15 P O BOX 126 304 26...

Отзывы: