Product application, installation and service manual
ELIS PLZE
Ň
a. s.
Ultrasonic Flow Meter SONOELIS SE 4015 and SE 4025
Page 7 of 44
ELIS PLZE
Ň
a. s., Lu
č
ní 15, P. O. BOX 126, 304 26 Plze
ň
, Czech Republic, tel.: +420/377 517 711, fax: +420/377 517 722
Es 90 337 K/c
3.2.4 Electronic unit
The flow-meter electronic unit (see Fig. 5) is embedded in a plastic box with a steel sheet attached at the back for
vertical mounting. At the front panel on the box there are meter type designation and product name, production series
number, manufacturer’s name and logo, two-line back-lighted display unit and a membrane keyboard. At the bottom of
the box under a removable plastic cover are plastic grommets for cables of circular cross-section (one PG 9 and six or
seven PG 7 grommets). The grommets are intended for tight fitting of cables of external diameter 6 to 8mm (PG 9) and
4 to 6mm (PG 7). At the bottom of the box there is also an earthing bolt. Both the front panel and the terminal board
cover can be sealed. Instead of one PG 7 grommet it is possible to fit a four-pole connector for the RS 485
communication line.
IMPORTANT NOTICE: Prior to putting the meter in operation, check the proper tightening of all grommets with cables
and blinding of the unused ones.
IMPORTANT NOTICE: The electronic unit shall not be exposed to direct sunlight.
PUSH-BUTTONS
PLASTIC COVER OF
THE TERMINAL STRIPS
GROUNDING
DISPLAY
LOCKING
PIN
FRONT PANEL
MECHANISM
OFFICIAL SEALS
INSTALATION SEALS
SONOELIS
ELIS PLZEN a.s.
COMFORT
Fig. 5 Electronic unit – description
3.2.5 Flow meter handling
The arrows indicate
the correct manner
of the meter sensor handling. Lift the sensor body holding it by the
handling eyes screwed into both flanges (see Fig. 6).