ELIS PLZEŇ a. s.
Project design, installation and service manual
Page 28 o
f
33
Electromagnetic flowmeters of the type series
FLONEX FXx11x and FLONET FH30xx
Meter control instructions
ELIS PLZE
Ň a. s., Lucni 425/15, 301 00 Plzeň, Czech Republic tel.:+420 377 517 711, fax:+420 377 517 722
Es 90 686 K
5.1.4.13. Production setting
The menu item P
roduction setting
is not accessible to the user. Data and information intended for the
manufacturer only are protected by a password. This arangement concerns:
Sensor constant
ADC setting
5.1.4.14. Simulation mode
The commencement of simulation mode will be confirmed
b
by a short flash in menu item Simulation mode start
The completion of simulation mode will be confirmed
by a short flash in menu item Simulation mode end
Upon confirming the selection of item
Simulated flow rate
by
ENTER
, a window will open for setting the
value of simulated flow rate in litres per second. Regarding conversion of volume flow rate in various units
to flow rate in l/s, see the procedure for setting
100%
.
Confirm the set flow-rate value by
ENTER
whereby the display will return to the previous menu item
(
Simulated flow rate
).
Select the command
Simulated mode start
and confirm by
ENTER.
The commencement of the simulation
procedure will be confirmed by a short flash of the
Simulated mode start
field.
All outputs (analog, OUT1 and OUT2 outputs and the figure on the instantaneous flow rate line on the
display) will be set at the value corresponding to the simulated flow rate. Next to the value of the simulated
flow rate on line 1 of Basic screen will appear the text message
SIM
.
Return to Main menu:
1 x ESC
.
Upon return from Main screen to item
Simulated flow rate
, use button
ENTER
to discontinue the
simulation mode. Termination of the simulation mode will be confirmed by a short flash of the
Simulated
mode end
field.
All outputs (analog, OUT1 and OUT2 outputs and the figure on the instantaneous flow rate line on the
display) will return to the status existing prior to the commencement of the simulation mode procedure.
Simulated flow rate
Simul. mode start
Simul. mode end
Simulated flow rate
l/s
+1.00
ENTER
Simulated flow rate
Simul. mode start
Simul. mode end
ENTER
ENTER
► ▲
Simulated flow rate
Simul. mode start
Simul. mode end
ENTER
► ▲