ELGA PURELAB UHQ II & PS Service Manual
PURELAB UHQ service issue 2 - 03/07
7
2. PURELAB UHQ II & PS features
Function
UHQ II
UHQ PS
RO make-up
4.0 l/hr
N/A
RO Drain flow l/hr
66 l/hr
N/A
Recirculation flow (l/hr)
Up to 45 l/hr
Permeate quality sensor
No
N/A
DI quality sensor
Yes with built-in temperature compensation
UV treatment
Yes
Point of use filter option
No
Reservoirs
Integral
4 litres (approx.)
Product water quality
18.2 M
Ω
@ 25
°
C
TOC (typical)
<20 ppb
Bacteria (see note)
<1 CFU/ml
Particles
0.05
µ
m
Note. Bacteria performance depends on correct operation and maintenance in accordance
with the operator manual.