PURELAB Prima 7/15/30
Operator Manual
ELGA
Page 12
PURELAB Prima 7/15/30 Version 01 09/08
Electrical
The units can be connected universally to any electrical supply
in the range of
100 - 240V and 50 - 60Hz.
The mains lead is
supplied with a molded plug on one end and a molded
connector to the unit on the other. The unit should be
connected to an earth.
Drain
A semi rigid flexible connection to a sink or suitable drain
capable of handling at least 1.5 l/min is required. The drain
point should have a gravity fall below the level of the unit and
any connections direct to drain should have an air-break device
fitted.
Feed Water
The feed water should be of good quality and comply with
specifications provided.
This should enter the unit via an
8mm
(5/16") OD
semi rigid tube, and should be in the temperature
range 1 to 35°C.
CAUTION!
Operating with feed water temperatures
outside the range from 1 to 35°C will cause
damage to the
PURELAB Prima
unit.
For pressurized feeds, the minimum direct inlet pressure is
4 bar (60 psi) (6 bar (90 psi) for
PURELAB Prima 30)
and
maximum inlet pressure is 6 bar (90 psi). Higher feedwater
pressures must be reduced using a pressure regulator valve
(Part No. LA 512).
Reservoir feeds to the
PURELAB Prima
unit should be
positioned at the same height, or above the unit, to provide a
positive flooded inlet pressure.
PURELAB Prima
Level control cable
RESERVOIR
Outlet
Inlet
Drain
All tube connections 8mm (5/16")
PURELAB Prima
Unit Installed with Storage Reservoir
Direction of
water flow
Collar
Collar
Mesh filter
Strainer
Feed Water Strainer
Mains power socket
Mains power lead
Fuse
ON/OFF
switch
Electrical Connections