PURELAB flex
Section 1 – Process Description
Issue 4
Page
6
of
45
Flex 2
Process Flow
1.
The pre-purified water enters the system through the inlet solenoid valve and flow sensor (FS1)
2.
It then passes through recirculation pump (P1), Quality sensor (QS1) and temperature sensor
(TS1)
3.
The water passed through the 185/254nm UV and quality sensor (QS3) before entering the
optional DI pack. Quality sensor (QS3) is used to monitor TOC levels.
4.
The water quality is then measured by a quality sensor (QS2) and a temperature sensor (TS2)
before reaching the dispense handset.
5.
When the system is not dispensing, the water will flow through the handset and a 15 psi non
return valve (V3) before returning to the recirculation pump.
6.
When the system is dispensing, valve (V2) will open and water will be directed out of the
dispense nozzle.
1
Inlet from
pre-purified
source
Flow Lines
P1
QS2
QS1
TS2
V1
V2
TS1
FS1
QS3
Notes:
V3
Optional
DI Pack