ELGA
PURELAB flex 1/2 - US Operators Manual
PURELAB flex 1/2 - US MANU40562 Version 02 – 10/15
Page 11
8.
TECHNICAL SPECIFICATIONS
FEEDWATER
Model
PURELAB flex
PURELAB flex
(with Purification
Pack)
PURELAB flex (with
UV, Purification
Pack and TOC)
SOURCE
Originally from potable supply, then pre treated.
Preferably reverse osmosis (RO) or filtered service
deionization (SDI) or distilled.
Note:- Mixed bed or twin bed deionized supplies should
be cation limited at exhaustion.
Conductivity
1
< 1
S/cm
Contaminant
Free Chlorine
< 0.05 ppm
Carbon Dioxide
<0.1 ppm
Silica
< 2 ppm
Fouling Index
1
Particulates
A 0.2 micron membrane pre filter is recommended for all
non-RO feeds to extend point-of-use filter life.
Iron/Manganese
< 0.1 ppm Fe/Mn
Organics (TOC)
N/A
N/A
<20 ppb
TEMPERATURE
4 - 40ºC (Recommended 15 - 25ºC)
FLOWRATE
(maximum requirement at
15ºC)
>2 l/min
Drain Requirements (gravity fall with air
gap)
None required
1
If feed water is in the range of 1
S/cm- 30
S/cm you will need a pre-conditioning
cartridge.
FEEDWATER PRESSURE
Maximum pressure
22psi (1.5bar)
Minimum pressure
Flooded suction
DIMENSIONS and WEIGHTS
Dimensions
Width 236mm, Depth 374mm, Height minimum 900mm,
Height maximum 1020mm.
Model
PURELAB flex
PURELAB flex
(with Purification
Pack)
PURELAB flex (with
UV, Purification
Pack and TOC)
Operational weight
10kg (22lb)
10.5 kg (23.1lb)
11kg (24.2lb)
Installation
Bench / Wall