Micra MANU39752 Version 01 – 02/12
10
. TECHNICAL SPECIFICATIONS
1
Purification Pack life may be low with feedwater >1400μS/cm
FEEDWATER
Model
Micra
SOURCE
Potable tap water as detailed below.
Conductivity
< 2000µS/cm
1
Hardness
<250ppm as CaCO
3
Free Chlorine
<0.5ppm Cl
2
Silica
<30ppm SiO
2
Fouling Index
<10
Iron/Manganese
< 0.1ppm Fe/Mn
Organics (Total Organic Carbon)
<3ppm TOC
TEMPERATURE
4-40°C (recommended 15-25°C)
FLOWRATE
(maximum requirement at 15ºC)
Up to 72L/h
FEEDWATER PRESSURE
Minimum Inlet Pressure
1bar (15psi)
Maximum Inlet Pressure
2bar (30psi)
2
2
In the situation that the Inlet Pressure >2bar (30psi) add a pressure regulator
CONNECTIONS
Inlet
1/4" OD tube
Outlet
1/4" OD tube
Drain
1/4" OD tube
DIMENSIONS and WEIGHTS
Dimensions
Width 22.5”, Depth 10.3”, Height 15”
Width 505mm, Depth 230mm, Height 335mm
Operational weight
26.5lb
12kg
Installation
Floor/ Bench / Wall
ELECTRICAL REQUIREMENTS
Mains input
100-240Vac, 50-60Hz
System control voltage
(not including pumps and UV)
24Vdc
Power consumption (peak demand)
120VA