Page 13

Micra MANU39752 Version 01 – 02/12

8

SELF HELP GUIDELINES

This section covers some of the issues that could occur with your Micra, and provides the troubleshooting 
steps to help you to resolve the issue yourself.

When troubleshooting doesn’t resolve your issue, please call your local ELGA LabWater representative 
(See Section 8 - Useful Contact Details).

 

   

WARNING! 

 ALWAYS ENSURE THAT THE ELECTRICAL POWER SUPPLY IS 
ISOLATED BEFORE WORKING INSIDE Micra.

DISPLAY

RECOMMENDED ACTION

NO LIGHT

Check power supply connection, turn the system on. 
If the system does not react, contact technical support. 

GREEN LIGHT ON

None – the system is working correctly.

GREEN and RED LIGHT ON 

The water quality is below the setpoint. 

Change the Purification Cartridge, 10 Micron Depth Filter and 

Carbon Filter.

RED LIGHT ON - continuous

Check if the Purification Cartridge is fitted in correctly

RED LIGHT ON – flashing

Purification Cartridge/10 Micron Depth Filter/Carbon Filter 

nearing expiry date.

Содержание Micra

Страница 1: ...Micra Operator Manual ...

Страница 2: ...LTH AND SAFETY NOTES 14 9 1 Environment 14 9 2 Electricity 14 9 3 Residual Biostat 14 9 4 Control of Substances Hazardous to Health COSHH 14 10 TECHNICAL SPECIFICATIONS 15 11 USEFUL CONTACT DETAILS 17 12 WARRANTY CONDITIONS OF SALE 18 The information contained in this document is the property of VWS UK Ltd trading as ELGA LabWater and is supplied without liability for errors or omissions No part o...

Страница 3: ...r requirements 1 3 Environment The Micra unit should be installed on a flat level surface in a clean dry environment The unit can also be wall mounted against a vertical wall capable of supporting the weight of the unit 1 4 Customer Support If you need help with your Micra please call your local ELGA representative For the address of the nearest ELGA LabWater Sales and Service office visit the cou...

Страница 4: ...2 Version 01 02 12 ELGA Micra Operators Manual 2 YOUR GUIDE TO THE MICRA On Off Button Connections 10 Micron Depth Filter Carbon Filter RO Module Purification Cartridge 3 Outlet Purified Water 2 Drain 1 Inlet Potable Water ...

Страница 5: ...mounted floor mounted or bench mounted 4 1 Level Control It is recommended that Micra is connected to a suitable reservoir and operated in an automatic mode using the level control supplied 1 PLUG in the level control POSITION the level control in the reservoir as pictured below Max water level Min water level CONSUMABLES Part No Description Typical Service Life Max Shelf Life LC226 Consumable kit...

Страница 6: ... 5 Micra MANU39752 Version 01 02 12 4 2 Power Cable Install 1 UNSCREW the housing for the Depth Filter Always use only the power cable provided with the unit 2 UNSCREW the nuts securing the protective plate 3 SLIDE in the power cable ...

Страница 7: ...TTACH the plate back on its original position PLUG in the lead 4 3 10 Micron Depth Filter and Carbon Filter Installation 1 UNPACK the 10 Micron Depth Filter Carbon Filter 10 Micron Depth Filter 2 PLACE the 10 Micron Depth Filter inside the housing and SCREW it back in place ...

Страница 8: ...om its packaging PLACE the Carbon Filter inside the housing and SCREW it back in place 4 4 DI Pack installation 1 REMOVE a new Purification Cartridge from its packaging REMOVE the sealing plugs from the inlet and outlet ports WET O rings and SLIDE the cartridge into the position POSITION cartridge onto spigots at rear and PUSH into unit 1 2 ...

Страница 9: ...o rinse the system Once completed the system is ready to use CONNECT the power 4 5 Water Connections Follow instruction below when connecting Portable water inlet Purified water outlet and Drain outlet UNSCREW compression nut REMOVE transit fitting INSERT 1 4 tubing HAND TIGHTEN compression nut 1 2 TURN to extra tighten NOTE At the Portable water inlet ensure pressure will not exceed 2 bar 30 psi ...

Страница 10: ...of water is available 5 2 Manual operation No Reservoir Controls PRESS process button to start green light will appear PRESS process button to stop no light on the display 6 SETTING WATER QUALITY ALARM The water quality alarm is automatically set up to warn you when the water quality is below 10MΩ cm It can be adjusted to 5MΩ cm or 2MΩ cm by following this procedure WARNING PRIOR TO ANY MAINTENANC...

Страница 11: ...ANU39752 Version 01 02 12 2 REMOVE the top cover 3 According to your needs MOVE the link from the default position 1 Link Positioning Water Quality Alarm Set point MΩ cm 13 and 14 10 7 and 8 2 5 and 6 5 2 1 Link in default position 13 1 14 2 TH2 ...

Страница 12: ...10 Micron Depth Filter and Carbon Filter are exhausted These cartridges should be always exchanged at the same time to ensure optimal results 7 1 Replacing the Reverse Osmosis Module LC229 Micra is fitted with a Reverse Osmosis RO module The RO Module may need replacing if the permeate water purity or flow rate is not adequate and does not meet predicted or previous performance For information reg...

Страница 13: ...ations is provided with the tablets and should be read before the tablet is used 1 TURN system off 2 UNSCREW carbon filter housing then remove carbon filter 3 ADD CS1 tablets to housing then allow to dissolve 4 REPLACE the filter housing Do not install a new carbon filter 5 DIRECT Purified Water outlet to Drain 6 TURN on system for 1 hour 7 TURN off system EXCHANGE the Filters and the Purification...

Страница 14: ...S ENSURE THAT THE ELECTRICAL POWER SUPPLY IS ISOLATED BEFORE WORKING INSIDE Micra DISPLAY RECOMMENDED ACTION NO LIGHT Check power supply connection turn the system on If the system does not react contact technical support GREEN LIGHT ON None the system is working correctly GREEN and RED LIGHT ON The water quality is below the setpoint Change the Purification Cartridge 10 Micron Depth Filter and Ca...

Страница 15: ...nmental specification could result in damage to the system Micra is not designed for use in fume cupboards where chemicals could damage the system WARNING ONLY USE THE APPLIANCE COUPLER MAINS LEAD AND POWER SUPPLY PROVIDED THE USE OF THESE WILL ENSURE ADEQUATE EARTH PROTECTION IS PROVIDED IF THE EQUIPMENT IS USED IN A MANNER NOT SPECIFIED BY ELGA THE PROTECTION PROVIDED BY THE EQUIPMENT MAY BE IMP...

Страница 16: ... C recommended 15 25 C FLOWRATE maximum requirement at 15ºC Up to 72L h FEEDWATER PRESSURE Minimum Inlet Pressure 1bar 15psi Maximum Inlet Pressure 2bar 30psi 2 2 In the situation that the Inlet Pressure 2bar 30psi add a pressure regulator CONNECTIONS Inlet 1 4 OD tube Outlet 1 4 OD tube Drain 1 4 OD tube DIMENSIONS and WEIGHTS Dimensions Width 22 5 Depth 10 3 Height 15 Width 505mm Depth 230mm Hei...

Страница 17: ...olicy of continual improvement we reserve the right to alter the specifications given in this document PRODUCT WATER SPECIFICATION Volume of Pure Type 2 Water used per working day Up to 25L day Pure water delivery flow rate Temperature 15 C Up to 8L h Inorganics Resistivity 25 C 1MΩ cm ...

Страница 18: ...rations Centre Lane End Industrial Park High Wycombe Bucks HP14 3BY UK Tel 44 0 1494 887500 Fax 44 0 1494 887505 E mail info elgalabwater com For the address of the nearest ELGA LabWater Sales and Service office visit the country list on our website http www elgalabwater com Or contact ELGA LabWater at the number above ...

Страница 19: ...SE The warranty provided herein and the data specifications and descriptions of the VWS UK Ltd systems appearing in VWS UK Ltd published catalogues and product literature may not be altered except by express written agreement signed by an officer VWS UK Ltd Representations oral or written which are inconsistent with this warranty or such publications are not authorised and if given should not be r...

Страница 20: ......

Отзывы: