Micra Operators Manual
Page 9
Micra MANU39752 Version 01 – 02/12
5
. OPERATION
Micra provides CLSI Type II pure water for manual and semi-automatic clinical analyzers with daily
consumption up to 25L.
5.1 Automatic operation (Reservoir Controls).
ENSURE the unit is connected to a suitable reservoir and the level control is installed.
PRESS process button.
The unit will start filling the tank automatically, ensure that a reservoir of water is available.
5.2 Manual operation (No Reservoir Controls).
PRESS process button to start (green light will appear).
PRESS process button to stop (no light on the display).
6
.
SETTING WATER QUALITY ALARM
The water quality alarm is automatically set up to warn you when the water quality is below 10MΩ/cm.
It can be adjusted to 5MΩ/cm or 2MΩ/cm by following this procedure:
WARNING!
PRIOR TO ANY MAINTENANCE, TURN OFF MICRA
AND DISCONNECT IT FROM ELECTRICAL POWER SUPPLY.
WARNING!
IN CASE OF UNEXPECTED POWER SHORTAGE, THE UNIT WILL
RESUME WORKING AUTOMATICALLY, WITH 1 MIN FLUSH IN
THE BEGINNING.
1) UNSCREW the screws securing the top cover.