7. OPERATION
The
MEDICA EDI 15/30
will function automatically and signal alarm
conditions to ensure efficient system management and corrective
action.
Note:
To ensure water purity is maintained it is important to
leave the system in process mode.
7.1 Dispensing
The system should be in process mode with no alarms activated.
If the system is not processing, press PROCESS to start. Wait until
the water reaches the preferred purity level before use. Dispense
water from the dispense tap by pulling the tap handle forward.
Note:
It will not be possible to dispense from the tap while supplying
water to the analyzer.
7.2 Intermittent Mode
During periods of non-use the system will automatically
operate in intermittent mode to maintain water quality. This
mode will function after the reservoir has been filled and the
level maintained for 60 minutes. The system will then
recirculate the reservoir contents for 5 minutes every 30
minutes.
During demand from the Analyzer, re-circulation will be
initiated.
7.3
Alarm Conditions
See section 11