MEDICA EDI 15/30 

Operator Manual 

ELGA 

Page 18 

MEDICA EDI 15/30

 Version 1 05/16 

Step 8 - Product water temperature alarm settings TS2 

This  screen  is  used  to  select  the  value  at  which  the  product 
water temperature alarm will activate.  The alarm does not stop 
the  system  and  will  automatically  reset  if  the  temperature 
returns below the set point.  

1. 

PRESS 

  to  select  alarm  point  (increments  of  1, 

ranging from 20 to 50

C) 

 

OR

 

PRESS 

 to proceed to step 9. 

2. 

PRESS 

✔ 

to confirm selection. 

Step 9 - Auto-restart 

This allows selection of the AUTO restart option. If auto restart 
is  selected  the  system  will  automatically  restart  after  a  power 
failure.  In  manual  mode  the  system  will  stay  in  standby  after  a 
power failure. 

2. 

PRESS 

 to change mode (

 = ON) 

 

OR

 

PRESS 

 to proceed to step 18. 

2. 

PRESS 

 to confirm selection. 

Step 10 

– CVF replacement alarm setting 

This allows confirmation of when to sound the alarm to indicate 
recommended  replacement  of  the  LC136  CVF.  The  default 
setting is 182 days. 

1. 

PRESS 

 to confirm selection. 

Step 11 

– UV replacement alarm setting 

This allows confirmation of when to sound the alarm to indicate 
recommended replacement of the UV lamp. The default setting 
is 364 days. 

1. 

PRESS 

 to confirm selection. 

 

 

 

 

 

 

 

 

 

Содержание Medica EDI 15

Страница 1: ...Medica EDI 15 30 Operator Manual...

Страница 2: ......

Страница 3: ...rmission from VWS UK Ltd The copyright and all restrictions on reproduction and use apply to all media in which this information may be placed VWS UK Ltd pursue a policy of continual product improveme...

Страница 4: ...ng the external Pre treatment Assembly 11 6 4 Connecting the MEDICA EDI 15 30 13 6 5 Pre Start Up 15 6 6 Initial Controller Set Up 16 6 7 Setting Display and Consumable Replacement Reminders 20 6 8 Co...

Страница 5: ...he World 34 9 3 Sanitization Procedure for the recirculation loop and docking vessel reservoir 35 10 EMERGENCY BY PASS 36 11 TROUBLE SHOOTING 38 12 KEY TO CONTROL PANEL 39 12 1 Icons 39 12 2 User Alar...

Страница 6: ......

Страница 7: ...to the instructions in this document then the safety of the user may be compromised 1 2 Customer Support Service support and consumable items are available from your local supplier or distributor Ref...

Страница 8: ...is isolated before any items are changed or maintenance work performed The ON OFF switch is located on the left hand side of the system The mains power lead is located just behind the ON OFF switch W...

Страница 9: ...ant safety guidance is included in this document Please refer to the manufacturer for safety data sheets EfferSan is EPA registered as a sterilant high level disinfectant and sanitizer EfferSan is aci...

Страница 10: ...posite Vent Filter 6 months 2 years Life is an estimate only and will depend on the application and feed water quality Care should be taken to order the correct consumable items Replacement frequency...

Страница 11: ...rified water to clinical analyzers requiring a pressurized feed The MEDICA EDI 15 30 can be bench or wall mounted and an optional wall mounting kit is available A range of accessories are available to...

Страница 12: ...s the conductivity of the water The permeate water passes through the conditioning chamber of the resin cartridge pack which removes traces of hardness remaining in the water The water then passes thr...

Страница 13: ...level of water in the reservoir is below 10 or below 40 on start up the re circulation mode will be disabled until a level of 40 is reached The re circulation will restart automatically When the make...

Страница 14: ...dow and three program function control buttons Details of how to use the controls are provided in the appropriate sections The MEDICA EDI 15 30 control panel has a range of control icons as follows BU...

Страница 15: ...The system can also be wall mounted against a vertical wall capable of supporting the weight for this we recommend the use of the wall mounting kit Part No LA 610 CAUTION If system is to be wall mount...

Страница 16: ...e fitted WARNING IN COMMON WITH OTHER ELECTRO DEIONIZATION SYSTEMS THIS SYSTEM PRODUCES SMALL AMOUNTS OF HYDROGEN IN THE DRAIN LINE IF THIS IS ALLOWED TO BUILD UP IN A RESTRICTED ENVIRONMENT THIS COUL...

Страница 17: ...d WARNING THIS ACCESSORY IS SUITABLE FOR MOUNTING TO MOST TYPES OF MASONRY OR CONCRETE TYPE WALLS 2 Using the wall mount plate as a template spot drill the four mounting holes into the wall 3 Remove t...

Страница 18: ...on Valve is in the CLOSED position 5 If feedwater tube is 3 8 OD CONNECT directly to inlet port of Isolation Valve 6 If feedwater tube is 5 16 8mm OD INSTALL 3 8 5 16 Stem Reducer before connecting 7...

Страница 19: ...on loop Step 1 Fitting Tubes 1 PUSH in collet on connector 2 PULL out transit plug 3 CUT a clean square end on an 8mm 5 16 OD semi rigid drain tube 4 PUSH tube into connector CAUTION Do not restrict d...

Страница 20: ...al scheme is illustrated below as are the recirculation delivery pump connections Electrical Connections Reservoir Level Connections MEDICA EDI 15 30 system installed with Docking Vessel Reservoir RES...

Страница 21: ...nlet system pressure if necessary The MEDICA EDI 15 30 will operate on an inlet pressure between 0 bar and 2 0 bar 30 psi with the water flowing 4 CHECK all hydraulic connections are water tight and t...

Страница 22: ...ce of the MENU icon on the control screen 2 PRESS the MENU button to go to the next screen to activate the set up menu sequence 3 A series of set up screens will now be displayed Various control icons...

Страница 23: ...rity to be set to either M cm or S cm This only refers to the quality measurement in the recirculation loop 1 PRESS SCROLL to change mode M cm or S cm OR PRESS to proceed to step 6 2 PRESS to confirm...

Страница 24: ...s allows selection of the AUTO restart option If auto restart is selected the system will automatically restart after a power failure In manual mode the system will stay in standby after a power failu...

Страница 25: ...alarm setting This allows confirmation of when to sound the alarm to indicate recommended replacement of the LC125 filter The default setting is 182 days 1 PRESS to confirm selection Step 14 Sanitizat...

Страница 26: ...date and proceed to Step 2 2 PRESS to confirm that resetting is required OR PRESS to abort reset 3 PRESS Step 2 UV Lamp LC105 replacement date Setting this screen will cause the UV Lamp timer to reset...

Страница 27: ...der date proceed to operation 2 OR PRESS to accept reminder date and complete the settings procedure 2 PRESS confirm that resetting is required OR PRESS to abort reset 3 PRESS Accessing the Process On...

Страница 28: ...d fittings 1 CHECK the water supply is turned on PRESS the PROCESS button and the system will start a 30 minute flush sequence where water will be directed to drain through a flush valve 2 LEAVE the s...

Страница 29: ...ntil the water reaches the preferred purity level before use Dispense water from the dispense tap by pulling the tap handle forward Note It will not be possible to dispense from the tap while supplyin...

Страница 30: ...WARNING ALWAYS CHECK THAT THE MAINS ELECTRICAL POWER AND FEED WATER SUPPLIES ARE SWITCHED OFF BEFORE ATTEMPTING TO REPLACE THE MEDICA EDI 15 30 CONSUMABLES Frequency of Consumable Replacement The fol...

Страница 31: ...1 PRESS both red pressure relief buttons located on top of the pre treatment housings to relieve residual pressure from the system 2 UNSCREW each 20 filter bowl using the bowl removal spanner supplie...

Страница 32: ...an absorbent cloth to soak up the water and RE TIGHTEN the bleed nipple WARNING ENSURE THE SYSTEM IS ISOLATED BEFORE REMOVING THE CARTRIDGE PACK Step 2 Remove Cartridge Pack 1 OPEN the front door 2 PU...

Страница 33: ...prevent splashing of internal components 12 CLOSE the upper bleed nipple on the 0 2 m capsule filter 13 After 5 minutes PRESS the PROCESS button to stop flow 14 OPEN V1 and CLOSE V2 to re direct flow...

Страница 34: ...Hold on to the lamp pins in case the lamp falls out and breaks Step 3 Remove UV Lamp LC105 1 REMOVE the old UV lamp from the center bore of the housing and discard Note The consumable is non hazardous...

Страница 35: ...Step 2 Remove 0 2 m capsule filter 1 OPEN the front door panel 2 IDENTIFY the 0 2 m capsule filter LC125 3 REMOVE the filter from its retaining clips and pull forward 4 REMOVE the 6mm clear tube from...

Страница 36: ...and set aside to gain access to the inlet strainer 5 DEPRESS collars on both sides of strainer elbows and disconnect tubing 6 REMOVE the inlet strainer from its position Step 2 Dismantle the Inlet St...

Страница 37: ...n Strainer 1 REPOSITION the re circulation strainer 2 REFIT tubes to re circulation strainer ENSURE it is facing the correct direction 3 REPOSITION the conditioning cartridge 4 REPOSITION the side pan...

Страница 38: ...mation They will also advise on procedures before and after extended periods of non use 9 1 Sanitization of RO Modules for use within the United States WARNING ALWAYS WEAR RUBBER GLOVES APRON AND FACE...

Страница 39: ...rt 1 CUT 1 EfferSan tablet into quarters using tablet cutter 2 UNSCREW cap on sanitization port 3 ADD of EfferSan tablet CAUTION Ensure the correct dosage of chemical is used 4 REFIT cap on sanitizati...

Страница 40: ...8 2 for details 8 INSTALL the LC213 sanitization bypass cartridge pack follow LC212 procedure See Section 8 2 for details 9 TURN ON the electrical supply The system will display the sanitization stan...

Страница 41: ...and docking vessel reservoir is recommended to be performed if high levels of bacterial contamination are found in the product water feeding the analyser Should bacterial contamination of the product...

Страница 42: ...ge cartridge pack LC212 See Section 8 2 3 OPEN the sanitization isolating valve V2 see page 12 for details 4 CLOSE the analyser isolating valve V1 see page 12 for details Step 3 Re establish supply 1...

Страница 43: ...ctivity 605 S cm Water consumption of analyzer 20 l hr Time between pack changes 29 400 605 x 20 Time between Pack changes 2 4 hrs In the absence of information on mains water conductivity assume a pa...

Страница 44: ...see Section 8 2 Replacing the LC212 cartridge Pre treatment cartridge replacement alarm Replace pre treatment filters see Section 8 1 Replacing the pre treatment filters 0 2 m capsule filter replaceme...

Страница 45: ...ON ICON DESCRIPTION Accept Menu Auto restart Mute alarm Scroll back Option OFF Step back Option ON Bell Output Standby Recirculation Calibration point Reset Cancel Right Clock Sanitization reminder Co...

Страница 46: ...in place 67 LC213 Bypass Cartridge pack has expired 69 Replace Pre treatment Filters 70 Replace UV Lamp 71 Replace 0 2 m Capsule Filter 72 Replace Composite Vent Filter 74 Sanitization Reminder 78 Con...

Страница 47: ...irement 100 l hr Drain Requirements gravity fall with air gap Maximum during Service 70 l hr Feedwater Pressure Maximum 8 0 bar 120 psi Minimum Flooded Suction Pre Treated Water Specification MEDICA E...

Страница 48: ...tart after power failure Audible alarm Water purity units Water purity Selectable Selectable Micro Siemens cm or Megaohm cm Alarm set points Indicators Reverse osmosis permeate water De ionized water...

Страница 49: ...ation In line 0 2 m Filtration Purified Water Specification MEDICA EDI 15 MEDICA EDI 30 Make up rate 15l hr 30 l hr Daily output nominal max Up to 300 Litres 24 hour day Dispense rate from tap 1 3 l m...

Страница 50: ...30 Version 1 05 16 MEDICA EDI 15 30 Reverse Osmosis RO Capacity Charts Graph 1 RO Production Rate vs Feed Water Temperature for MEDICA EDI 15 30 0 5 10 15 20 25 30 35 40 45 50 0 5 10 15 20 25 30 35 Fl...

Страница 51: ...VWS UK Ltd promptly of any such defect The exclusive remedy provided herein shall not be deemed to have failed of its essential purpose so long as VWS UK Ltd is willing and able to repair or replace...

Страница 52: ...customer from the use of its process systems Products or components manufactured by companies other than VWS UK Ltd or its affiliates Non VWS UK Ltd products are covered by the warranty if any extend...

Страница 53: ...Water Lane End Industrial Park High Wycombe Bucks HP14 3BY UK Tel 44 0 203 567 7300 Fax 44 0 203 567 7305 E mail info elgalabwater com For the address of the nearest ELGA LabWater Sales and Service of...

Отзывы: