MEDICA EDI 15/30
Operator Manual
ELGA
Page 18
MEDICA EDI 15/30
Version 1 05/16
Step 8 - Product water temperature alarm settings TS2
This screen is used to select the value at which the product
water temperature alarm will activate. The alarm does not stop
the system and will automatically reset if the temperature
returns below the set point.
1.
PRESS
to select alarm point (increments of 1,
ranging from 20 to 50
C)
OR
PRESS
✔
to proceed to step 9.
2.
PRESS
✔
to confirm selection.
Step 9 - Auto-restart
This allows selection of the AUTO restart option. If auto restart
is selected the system will automatically restart after a power
failure. In manual mode the system will stay in standby after a
power failure.
2.
PRESS
to change mode (
= ON)
OR
PRESS
✔
to proceed to step 18.
2.
PRESS
✔
to confirm selection.
Step 10
– CVF replacement alarm setting
This allows confirmation of when to sound the alarm to indicate
recommended replacement of the LC136 CVF. The default
setting is 182 days.
1.
PRESS
✔
to confirm selection.
Step 11
– UV replacement alarm setting
This allows confirmation of when to sound the alarm to indicate
recommended replacement of the UV lamp. The default setting
is 364 days.
1.
PRESS
✔
to confirm selection.