25
elements
enabling technologies for Life Science
eONE - User’s guide
Figure 21
eONE Controls - Base version
Finally, below are listed the controls that are present only in the eONE Base ver-
sion, which has some limitations compared to the VP and HS versions:
Triangular Wave check box
Vhold tab
The
Triangular Wave
control is used to force a 50 Hz triangular wave control volt-
age of peak to peak amplitude of 200 mVpp. This voltage signal is commonly used
to control lipid bilayer formation and to estimate the input capacitances, as de-
scribed in the “Statistic Data” section.
The
Vhold
box in the Vhold tab enables the user to select value of the constant
holding potential applied at the eONE input pin in the following range:
from -384 mV to +384 mV with 1mV steps