eONE - User’s guide
Figure 14
eONE Controls
–
Voltage Protocols
Constant Voltage protocol
Figure 15
eONE Controls
–
Voltage Protocols
Triangular wave protocol
Figure 16
eONE Controls
–
Voltage Protocols
Seal test protocol
Figure 17
eONE Controls
–
Voltage Protocols
Conductance test protocol
Constant voltage
protocol:
The constant voltage is considered as a protocol with only the
Vhold
parame-
ter.
Triangular wave
protocol:
It can be used to control the bilayer formation and it is used for the Capaci-
tance estimation feature (see details section 3.1.7).
Seal test
protocol:
It is commonly used for setting up the sealing of the pipette to the cell in
patch clamp studies. Usually the
Ne
parameter is set to 0 to repeatedly apply
the stimulus to the cell.
Conductance test
protocol:
It can be used to semi-automatically perform a channel’s conductance char-
acterization.