NM20216A-F
15
Technical data
1
1.3.7.
EEG main electronics
• Signal acquisition module (SAM) for 12 channels
• one digital signal processor (DSP) per SAM
• Resolution of A/D conversion: 16 bits
• Type of coupling: dc
• Input stage realized with instrumentation amplifiers with software
controlled gain of 1 or 10, corresponding to SAM input ranges of
±
10 V or
±
1 V, respectively
• All conversions started simultaneously, synchronized with MEG
channels
• Sampling rate equivalent to that of MEG channels
• Low-pass filters identical to those in MEG channels (DSP-based)
• 2 SAM / remote processor unit (RPU)
• Connected over dedicated Ethernet to parallel real-time data acquisi-
tion computers. Accessible also through workstations.
• SAM boards reside in the EEG rack inside themain electronics cabinet
• Number of subracks: 1
• Number of SAM boards: 6 (64 ch.) / 12 (128 ch.)
• Powering: Separate power supply for each set of four slots inside the
main electronics cabinet
• Fan unit: mains operated.
1.4.
MEG and EEG Power Supplies
1.4.1.
Front-end MEG power supply
• Power rating: 400 W
• Mains: 230 V~, 47…63 Hz, from a dedicated isolation transformer
• Control: manual power switch
• Number of units: 1
• Output:
±
15,5 V
=
9 A (analog), 24V~ 4 A (for isolated EEG), +6,3 V
=
8 A (digital)
• Output cabling to the filter unit
• Primary fuses: 3.15 A slow blow
• Feeds also the non-isolated part of the EEG system
• Overload protection: current-limiting circuit
• Location: inside main electronics cabinet
Содержание Neuromag
Страница 1: ...Elekta Neuromag System Hardware Technical manual Revision F September 2005 Elekta Neuromag...
Страница 6: ...6 NM20216A F...
Страница 32: ...32 NM20216A F 2 Dimensions and weights Figure 2 2 Dewar front view Dimensions in millimeters 1100 630...
Страница 52: ......
Страница 53: ...NM20216A F 53 Appendix Additional site specific data sheets if applicable...