IMPORTANT!
• Do not apply blood on the
Check-strip
.
• Protect
Check-strips
against physical damage and keep them in a
safe
place.
If a
Check-strip
is damaged or lost, contact your local distributor for
advice.
Checking with CareSens control solution
Press the
OK
button and keep it pressed until
“CHK”
is displayed, then release the
OK
button.
A flashing strip icon will be displayed,
indicating that the meter is waiting for strip
insertion. Insert an
AutoSense Test
strip and
push it gently into the end position. Compare
the displayed code with code printed on the
vial that held the strip. If the codes differ,
insert another strip.